Non-invasive Intracranial Pressure Measurement Methods
   HOME

TheInfoList



OR:

Increased
intracranial pressure Intracranial pressure (ICP) is the pressure exerted by fluids such as cerebrospinal fluid (CSF) inside the skull and on the brain tissue. ICP is measured in millimeters of mercury ( mmHg) and at rest, is normally 7–15 mmHg for a supine adu ...
(ICP) is one of the major causes of secondary
brain ischemia Brain ischemia is a condition in which there is insufficient bloodflow to the brain to meet metabolic demand. This leads to cerebral hypoxia, poor oxygen supply in the brain and may be temporary such as in transient ischemic attack or permanent i ...
that accompanies a variety of pathological conditions, most notably
traumatic brain injury A traumatic brain injury (TBI), also known as an intracranial injury, is an injury to the brain caused by an external force. TBI can be classified based on severity ranging from mild traumatic brain injury (mTBI/concussion) to severe traumati ...
(TBI),
stroke Stroke is a medical condition in which poor cerebral circulation, blood flow to a part of the brain causes cell death. There are two main types of stroke: brain ischemia, ischemic, due to lack of blood flow, and intracranial hemorrhage, hemor ...
s, and intracranial hemorrhages. It can cause complications such as vision impairment due to intracranial pressure ( VIIP), permanent neurological problems, reversible neurological problems,
seizures A seizure is a sudden, brief disruption of brain activity caused by abnormal, excessive, or synchronous neuronal firing. Depending on the regions of the brain involved, seizures can lead to changes in movement, sensation, behavior, awareness, o ...
, stroke, and death. However, aside from a few Level I trauma centers, ICP monitoring is rarely a part of the clinical management of patients with these conditions. The infrequency of ICP can be attributed to the invasive nature of the standard monitoring methods (which require insertion of an ICP sensor into the
brain ventricle In neuroanatomy, the ventricular system is a set of four interconnected cavities known as cerebral ventricles in the brain. Within each ventricle is a region of choroid plexus which produces the circulating cerebrospinal fluid (CSF). The ventricu ...
or parenchymal tissue). Additional risks presented to patients can include high costs associated with an ICP sensor's implantation procedure, and the limited access to trained personnel, e.g. a
neurosurgeon Neurosurgery or neurological surgery, known in common parlance as brain surgery, is the medical specialty that focuses on the surgical treatment or rehabilitation of disorders which affect any portion of the nervous system including the brain, ...
. Alternative, non-invasive measurement of intracranial pressure, non-invasive methods for estimating ICP have, as a result, been sought.


Correlation-based approaches

Many approaches to non-invasive ICP estimation are based on the idea that something in the human head's anatomical structure or in the intracranial and extracranial physiology ''correlates'' with ICP. ''"Correlation-based"'' approaches can reflect factors associated with ICP only with limited accuracy (expressed by
systematic error Observational error (or measurement error) is the difference between a measurement, measured value of a physical quantity, quantity and its unknown true value.Dodge, Y. (2003) ''The Oxford Dictionary of Statistical Terms'', OUP. Such errors are ...
) and precision (expressed by the standard deviation of random error). Measuring absolute ICP value is limited by the need for individual patient-specific calibration. Absolute ICP values in mmHg or other units are needed to determine the correct patient treatment. ''The only accurate, precise, and patient-specific, calibration-free, non-invasive, absolute ICP value measurement method relies not on the correlation, but on direct ICP and extracranial pressure comparison principle.''


Two depth transorbital doppler (TDTD)

An innovative method using a two-depth transorbital doppler (TDTD) of intracranial pressure quantitative absolute (ICP) value measurement relies on the same fundamental principle that is used to measure blood pressure with a
sphygmomanometer A sphygmomanometer ( ), also known as a blood pressure monitor, blood pressure machine, or blood pressure gauge, is a device used to measure blood pressure, composed of an inflatable cuff to collapse and then release the artery under the cuff i ...
. A sphygmomanometer works using a pressure balance principle - an air-filled pressure cuff wrapped around the arm compresses the
brachial artery The brachial artery is the major blood vessel of the (upper) arm. It is the continuation of the axillary artery beyond the lower margin of teres major muscle. It continues down the ventral surface of the arm until it reaches the cubital fossa ...
to a point where blood can no longer flow. Externally applied pressure is equal to systolic blood pressure in this case. The examiner slowly releases the air from the cuff and uses a
stethoscope The stethoscope is a medicine, medical device for auscultation, or listening to internal sounds of an animal or human body. It typically has a small disc-shaped resonator that is placed against the skin, with either one or two tubes connected t ...
to listen for the return of blood flow. At the pressure balance point, where the pressure in the cuff equals systolic artery pressure, a ‘whooshing’ noise can be heard as blood flows through the artery again. Pressure balance based a non-invasive blood pressure meter does not need a patient-specific calibration. The TDTD method uses
Doppler ultrasound Doppler ultrasonography is medical ultrasonography that employs the Doppler effect to perform imaging of the movement of tissues and body fluids (usually blood), and their relative velocity to the probe. By calculating the frequency shift of a ...
to translate the pressure balance principle of blood pressure measurement with a sphygmomanometer to the measurement of ICP. The ophthalmic artery (OA), a unique vessel with intracranial and extracranial segments, is used as a pressure sensor and as a natural pair of scales for absolute ICP value in mmHg or mmH2O measurement. Blood flow in the intracranial OA segment is affected by intracranial pressure, while flow in the extracranial (intraorbital) OA segment is influenced by the externally applied pressure (Pe) to the eyeball and orbital tissues. As with a sphygmomanometer, a special pressure cuff is used - in this case, to compress the tissues surrounding the eyeball and also intraorbital tissues surrounding the extracranial segment of OA. External pressure changes the characteristics of blood flowing from inside the skull cavity into the eye socket. In place of the stethoscope, a Doppler ultrasound beam measures the blood flow pulsations in intracranial and extracranial segments of the ophthalmic artery. The non-invasive ICP meter based on this method gradually increases the pressure over the eyeball and intraorbital tissues so that the blood flow pulsation parameters in two sections of the OA are equal. At this pressure balance point, the applied external pressure (Pe) equals the intracranial pressure (ICP). This measurement method eliminates the main limiting problem of all other non-successful approaches to non-invasive ICP measurement, primarily the individual patient calibration problem. Direct comparison of arterial blood pressure (ABP) and externally applied pressure is the basic arterial blood pressure measurement principle, which eliminates the need for individual calibration. The same calibration-free fundamental principle is used in the TDTD non-invasive ICP absolute value measurement method. The mean value of OA blood flow, its systolic and diastolic values, pulsatility and other indexes are almost the same in both OA segments in the point of balance when ICP equals Pe. As a result of that, all individual influential factors (ABP, cerebrovascular auto-regulation impairment, individual pathophysiological state of patience, individual diameter, and anatomy of OA, hydrodynamic resistance of eyeball vessels, etc.) do not influence the balance of ICP equaling Pe and, as a consequence, such natural “scales” do not need calibration. Ragauskas A. et al. have already published the statistically significant results of a prospective clinical study on assessment of the accuracy and precision of proposed non-invasive absolute ICP value measurement method. The study shows that proposed method is the only quantitative noninvasive ICP absolute value (mmHg) measurement method which does not need an individual patient-specific calibration. High accuracy, precision, sensitivity and specificity of the proposed method are fully acceptable for clinical practice and for very wide applications in neurology, transplantology, intensive care, sports medicine, aerospace medicine, and combat casualty care. This method was further developed by the Company ''Vittamed Ltd'' along with consortium partners in EU FP7 projects BrainSaf
Brainsafe
Brainsafe II, and TBIcare.


Ultrasound time of the flight techniques

The majority of patented methods for noninvasive monitoring of ICP are based on an assumption that changes in ICP affect the physical dimensions and/or acoustic properties of the
cranial vault The cranial vault is the space in the skull within the neurocranium, occupied by the brain. Development In humans, the cranial vault is imperfectly composed in newborns, to allow the large human head to pass through the birth canal. During bir ...
or intracranial structures ( dura, brain tissue, brain ventricles, and/or intracranial vessels). The common drawback of all these methods is that they measure only relative changes of ICP as referenced to a baseline measurement during which absolute ICP is known, i.e. the ultrasound readouts need to be calibrated on each subject against an invasive measurement. Ultrasound ‘time of the flight’ methods for non-invasive ICP monitoring have not been extensively validated and currently, the majority of them do not seem to be accurate enough for a routine clinical use. Their original formulations usually do not specify locations for the
transducer A transducer is a device that Energy transformation, converts energy from one form to another. Usually a transducer converts a signal in one form of energy to a signal in another. Transducers are often employed at the boundaries of automation, M ...
's placement and do not address how the intentional or accidental use of different locations and/or angles of the transducers will affect the reliability of ICP estimates. It has also remained unexplored how the measurements are affected by the presence of intracranial pathological masses on the path of the ultrasound wave, or by brain masses shifts.


Cranium diameter

Dimensions of the cranium or its structures are determined with the ultrasound “time-of-the-flight” technique that measures the transit time of an
ultrasound Ultrasound is sound with frequency, frequencies greater than 20 Hertz, kilohertz. This frequency is the approximate upper audible hearing range, limit of human hearing in healthy young adults. The physical principles of acoustic waves apply ...
wave and its (potentially multiple) echoes on their path through the
cranium The skull, or cranium, is typically a bony enclosure around the brain of a vertebrate. In some fish, and amphibians, the skull is of cartilage. The skull is at the head end of the vertebrate. In the human, the skull comprises two prominent ...
and calculates the corresponding distance(s) using known ultrasound propagation velocities in different tissues (e.g. bone, brain, or fluid). Unfortunately, a reproducible quantitative relationship between the diameter of the cranium and ICP could not be established because ICP-induced changes in the cranium diameter are very small compared to the inter-individual variability of
skull The skull, or cranium, is typically a bony enclosure around the brain of a vertebrate. In some fish, and amphibians, the skull is of cartilage. The skull is at the head end of the vertebrate. In the human, the skull comprises two prominent ...
sizes, shapes and thicknesses.


Thickness of the dura

The method claims that ICP can be inferred from the thickness of the dura mater that is estimated from interference echoes of an ultrasonic wave. The utility of the method was successfully confirmed on four healthy subjects and four patients with intracranial hypertension, but larger validation studies have never been conducted as the method failed to attract enough interest among clinicians. This method also needs a calibration to the individual patients.


Cerebral ventricle

Michaeli proposed that ICP should be inferred from the magnitude and shape of pulsations of the third ventricle synchronous with the
cardiac cycle The cardiac cycle is the performance of the heart, human heart from the beginning of one heartbeat to the beginning of the next. It consists of two periods: one during which the heart muscle relaxes and refills with blood, called diastole, fo ...
or respiration, where the pulsations are measured along the propagation axis of an ultrasound wave. The method so far has not been independently validated, and the author provides no exact data from which one could estimate the accuracy of the method. However, the discussion in the body of the patent document suggests that the method is able to distinguish among three ranges of ICP (<20, 20–40 and >40mmHg) but cannot provide an exact value of ICP within the range because of the impossibility of the calibration to the individual patient. Method and device for non-invasive (NI) ICP measurement according to inventions of David Michaeli MD, PhD, based to the TRA (tissue resonance analysis) have 2 options: (1) The qualitative method makes an evaluation of mild (10-20mm.Hg), moderate (20-40) and severe (above 40mmHg) ICP elevation. These method is using NI, long-term recording of ICP waves patterns, Like Lundsbergs ICP waves. (2) Quantitative ''measurement'' OF ICP WAVES with special ICP formulation in mm.Hg., Developed new device and method for calibration of ICP for each patient; se
patent description


Brain parenchyma

More recently, multivariate methods have been proposed that derive ICP by combining the transit times with measured acoustic impedance,
resonant frequency Resonance is a phenomenon that occurs when an object or system is subjected to an external force or vibration whose frequency matches a resonant frequency (or resonance frequency) of the system, defined as a frequency that generates a maximu ...
and ultrasound velocity, or with a dispersion of the ultrasound wave on its way through the brain parenchyma. Ultrasound ICP monitors based on the latter approach, which were developed at Vittamed Technologijos (Kaunas, Lithuania), have shown an impressive agreement with invasively measured ICP, with an average difference of only 2–3 mmHg in a small clinical population. However Vittamed Technologijos time-of-flight technologies were developed further for other applications of neuro-monitoring technologies (including cerebral Autoregulation and Cerebral Compliance).


Transcranial Doppler ultrasonography

The TCD measures the
velocity Velocity is a measurement of speed in a certain direction of motion. It is a fundamental concept in kinematics, the branch of classical mechanics that describes the motion of physical objects. Velocity is a vector (geometry), vector Physical q ...
of blood flow through the major intracranial vessels by emitting a high frequency (>2
MHz The hertz (symbol: Hz) is the unit of frequency in the International System of Units (SI), often described as being equivalent to one event (or cycle) per second. The hertz is an SI derived unit whose formal expression in terms of SI base u ...
) wave from an ultrasound probe and detecting a frequency shift between the incident and reflected wave which directly correlates with the speed of the blood (the so-called Doppler effect). The measurement is taken over the regions of the skull with thinner walls (temporal region, back of the head, or through the eye), as the bones strongly attenuate the transmission of the ultrasound at these frequencies. TCD is primarily a technique for diagnosing various intracranial vascular disorders such as
emboli An embolism is the lodging of an embolus, a blockage-causing piece of material, inside a blood vessel. The embolus may be a blood clot (thrombus), a fat globule (fat embolism), a bubble of air or other gas ( gas embolism), amniotic fluid (amnio ...
,
stenosis Stenosis () is the abnormal narrowing of a blood vessel or other tubular organ or structure such as foramina and canals. It is also sometimes called a stricture (as in urethral stricture). ''Stricture'' as a term is usually used when narrowing ...
, or
vasospasm Vasospasm refers to a condition in which an arterial spasm leads to vasoconstriction. This can lead to tissue ischemia (insufficient blood flow) and tissue death (necrosis). Along with physical resistance, vasospasm is a main cause of ischemi ...
, and can be used to identify patients who are at risk of developing
cerebral ischemia Brain ischemia is a condition in which there is insufficient bloodflow to the brain to meet metabolic demand. This leads to poor oxygen supply in the brain and may be temporary such as in transient ischemic attack or permanent in which there is ...
in early phases of
traumatic brain injury A traumatic brain injury (TBI), also known as an intracranial injury, is an injury to the brain caused by an external force. TBI can be classified based on severity ranging from mild traumatic brain injury (mTBI/concussion) to severe traumati ...
or
stroke Stroke is a medical condition in which poor cerebral circulation, blood flow to a part of the brain causes cell death. There are two main types of stroke: brain ischemia, ischemic, due to lack of blood flow, and intracranial hemorrhage, hemor ...
. ICP can be estimated from the TCD measurements because it impedes the blood flow and consequently decreases the velocity of blood flow. Besides the mean velocity, pulsatility index (which is the difference between peak systolic and end diastolic velocity, divided by mean flow velocity), a fraction of the cycle in
systole Systole ( ) is the part of the cardiac cycle during which some chambers of the heart contract after refilling with blood. Its contrasting phase is diastole, the relaxed phase of the cardiac cycle when the chambers of the heart are refilling ...
and slopes of the TCD waveforms have been correlated with ICP. The estimates are however insufficiently accurate with the margin of error of ±10 - 15 mmHg. ''Physiosonics, Inc.'' used transcranial Doppler ultrasound to measure ICP indirectly by assessing the elasticity of the biological material in a defined part of the brain. However, the elasticity in the brain is highly dependent on many other variable individual factors apart from ICP, including arterial blood pressure, state of cerebral blood flow auto-regulation, and the level of
edema Edema (American English), also spelled oedema (British English), and also known as fluid retention, swelling, dropsy and hydropsy, is the build-up of fluid in the body's tissue (biology), tissue. Most commonly, the legs or arms are affected. S ...
. Therefore, this approach would require calibration and expert positioning.


Skull bones

Methods from this group attempt to derive ICP from mechanical properties of the skull bones rather than of the intracranial content. The underlying assumption is similar to that of the ultrasound time of the flight techniques: that the skull is not completely rigid so that changes in ICP result in a small but measurable skull expansion which creates additional stress within the skull bones and modifies their mechanical properties. The transfer function is derived by applying a wide-band, low frequency (<100 Hz) mechanical excitation at one location on the skull (via a piezo-transducer or an impact hammer) and comparing its spectrum to that of a signal received at another location on the upper half of skull. It is proposed that the measurement to be self-calibrated by obtaining the frequency response spectrum from a point on the base of the skull of the same subject, which is assumed not to be affected by ICP, or alternatively, pre-calibrated on subjects with normal ICP. Other methods from this group vary this basic approach of Mick in different ways. In Sinha's method resonant frequency of the skull bones is determined first, then a
sinusoidal A sine wave, sinusoidal wave, or sinusoid (symbol: ∿) is a periodic wave whose waveform (shape) is the trigonometric sine function. In mechanics, as a linear motion over time, this is '' simple harmonic motion''; as rotation, it correspond ...
excitation at the resonant frequency is delivered through a piezo-transducer, and ICP is calculated directly from the phase difference between the excitatory signal and response detected with a second transducer. Yost and Cantrell divided the process into two steps. In the first step, changes in the
circumference In geometry, the circumference () is the perimeter of a circle or ellipse. The circumference is the arc length of the circle, as if it were opened up and straightened out to a line segment. More generally, the perimeter is the curve length arou ...
of the cranium are calculated from the phase difference between a sinusoidal excitatory signal, delivered with a piezo-transducer, and the response that is received at a distance with another piezotransducer. In the second step, changes in ICP are calculated as a product of the changes in the cranium circumference and the elasticity constant of the skull that has been determined earlier by causing known changes in ICP while measuring the cranium circumference. ''Luna Innovations Incorporated'' (NASDAQ: LUNA) developed EN-TACT system, an ultrasound device for monitoring compartment syndrome. The technology was claimed to have applications for raised intracranial pressure. Based on research from the NASA Ames research center, the company used ultrasound to measure skull diameter changes caused by ICP changes. None of the aforementioned methods has been properly validated in relevant clinical populations, and their accuracy is unknown. One may assume however that it would be comparable to the ultrasound time-of-the-flight methods, and thus insufficient for a routine clinical use. On the other hand, Mascarenhas et al. conducted a study using an empty skull with an attached strain gauge, where they induced variations in intracranial pressure (ICP) by inflating a balloon. They observed a linear correlation between micrometric skull bone expansion (measured in mV) and ICP, demonstrating that the Monro-Kellie doctrine was not applicable. As consequence of this research, a non-invasive system (brain4care, São Carlos, Brazil) was developed. This system was initially validated in living animal models and subsequently in patients with brain injuries. The system demonstrated a high-fidelity reproduction of ICP pulse slopes (ICP waveforms) by comparing noninvasive skull pulse waveforms with those obtained invasively through ventricular and parenchymal probes. The system is entirely noninvasive and safe, acquiring raw skull pulse waveform data and transmitting it to cloud-based processing. It returns real-time processed parameters, such as the P2/P1 ratio and time-to-peak, directly to clinicians. These parameters are correlated with intracranial compliance and, by extension, intracranial pressure. The brain4care system has been FDA cleared and is currently deployed in numerous healthcare facilities across Brazil and the United States, where it is used in the care and research of patients with various neurological and non-neurological conditions, regardless of disease severity.


Tympanic membrane displacement

Tympanic membrane displacement (TMD) technique, proposed nearly twenty years ago by Marchbanks exploits the effect of intracranial pressure on the acoustic reflex, i.e. a reflex contraction of the stapedius and tensor tympani muscles in response to a sound. Normally, vibrations of the
tympanic membrane In the anatomy of humans and various other tetrapods, the eardrum, also called the tympanic membrane or myringa, is a thin, cone-shaped membrane that separates the external ear from the middle ear. Its function is to transmit changes in pressur ...
(eardrum) elicited by acoustic stimuli are transmitted through the chain of
ossicles The ossicles (also called auditory ossicles) are three irregular bones in the middle ear of humans and other mammals, and are among the smallest bones in the human body. Although the term "ossicle" literally means "tiny bone" (from Latin ''ossi ...
(malleus, incus, and stapes) in the middle ear to the oval window of the
cochlea The cochlea is the part of the inner ear involved in hearing. It is a spiral-shaped cavity in the bony labyrinth, in humans making 2.75 turns around its axis, the modiolus (cochlea), modiolus. A core component of the cochlea is the organ of Cort ...
. Vibrations of the footplate of stapes transmit through the oval window to the
perilymph Perilymph is an extracellular fluid located within the inner ear. It is found within the scala tympani and scala vestibuli of the cochlea. The ionic composition of perilymph is comparable to that of plasma and cerebrospinal fluid. The major ca ...
, which in turn causes the
endolymph Endolymph is the fluid contained in the membranous labyrinth of the inner ear. The major cation in endolymph is potassium, with the values of sodium and potassium concentration in the endolymph being 0.91  mM and 154  mM, respectively. ...
, the
basilar membrane The basilar membrane is a stiff structural element within the cochlea of the inner ear which separates two liquid-filled tubes that run along the coil of the cochlea, the scala media and the scala tympani. The basilar membrane moves up and down ...
, and the organ of Corti to vibrate, activating ultimately the acoustic sensor cells, the inner hair cells of the organ of Corti. The transfer function of this complex mechanical system under physiological conditions is modulated by the action of two small muscles of the middle ear, the tensor tympani, and stapedius. The tensor tympani arises from the cartilaginous portion of the auditory tube and the osseous canal of the sphenoid and, having sharply bent over the extremity of the septum, attaches to the
manubrium The sternum (: sternums or sterna) or breastbone is a long flat bone located in the central part of the chest. It connects to the ribs via cartilage and forms the front of the rib cage, thus helping to protect the heart, human lung, lungs, and ma ...
of the
malleus The ''malleus'', or hammer, is a hammer-shaped small bone or ossicle of the middle ear. It connects with the incus, and is attached to the inner surface of the eardrum. The word is Latin for 'hammer' or 'mallet'. It transmits the sound vibra ...
(hammer); its contraction pulls the malleus medially, away from the tympanic membrane, which tenses the membrane. The
stapedius The stapedius is the smallest skeletal muscle in the human body. At just over one millimeter in length, its purpose is to stabilize the smallest bone in the body, the stapes or stirrup bone of the middle ear. Structure The stapedius emerges from ...
, which emerges from the posterior wall of the tympanic cavity of the middle ear and inserts into the neck of the stapes (stirrup), prevents excess movements of the stapes by pulling it away from the oval window. The action of either muscle, therefore, dampens vibrations of the ossicles and reduces the amplitude of transmitted sounds for up to 20 dB. The muscles normally contract in response to vocalization, jawing and loud external sounds, which is accompanied by a small but measurable displacement of the eardrum from its initial position. Because cerebrospinal fluid and perilymph communicate through the cochlear aqueduct, an increase in intracranial pressure is directly transmitted to the footplate of the stapes, changing its initial position and affecting thereby the direction and magnitude of the displacement of the
eardrum In the anatomy of humans and various other tetrapods, the eardrum, also called the tympanic membrane or myringa, is a thin, cone-shaped membrane that separates the external ear from the middle ear. Its function is to transmit changes in pres ...
in response to a sound. The displacement can be measured with common tympanometers used for impedance audiometry that are portable and relatively inexpensive and easy to use (particularly the modern, computerized tympanometers with fully automated measurement procedure). Inward displacement (negative peak pressure on an audiogram) is suggestive of high, and outward of normal or low ICP. The direction and magnitude of TMD, however, depend not only on the initial position of stapes but also on numerous other factors that affect the acoustic impedance (integrity of the eardrum, condition of the ossicles, patency of the Eustachian tube, pressure and eventual presence of fluid or other masses in the middle ear) or the strength of the acoustic reflex (physiological variability of the reflex threshold, functional integrity of the cochlear and facial nerves, degree of eventual sensory hearing loss). In addition, the assumption that the pressure of
perilymph Perilymph is an extracellular fluid located within the inner ear. It is found within the scala tympani and scala vestibuli of the cochlea. The ionic composition of perilymph is comparable to that of plasma and cerebrospinal fluid. The major ca ...
is equal to ICP does not hold if the patency of the cochlear aqueduct is compromised, which is often the case in elderly subjects. Accuracy of TMD estimates of ICP was found to be at the order of ±15mmHg, which is not sufficient for a reliable quantitative assessment of ICP in clinical practice. An interesting method that involves direct manipulations on the tympanic membrane rather than relying on the acoustic reflex was proposed as one of the embodiments of a US patent by Ragauskas. First, a measurement of the position of the tympanic membrane needs to be obtained while ICP is zero (denoted as the baseline position). Equalization of ICP to the atmospheric pressure according to the inventor can be achieved non-invasively by tilting the head up, or the measurement can be taken during a neurosurgical operation. Later on, ICP can be measured by exerting an external pressure to the tympanic membrane and applying simultaneously the same pressure onto the oval window and inner ear (e.g. through the Eustachian tube) until the eardrum is moved back to the baseline position, which will happen when the exerted external pressure equals ICP. No data is provided in the patent nor is available from other sources that could support the utility of the concept in clinical practice.


Otoacoustic emission

TMD fails to provide accurate estimates of ICP mostly because the
acoustic impedance Acoustic impedance and specific acoustic impedance are measures of the opposition that a system presents to the acoustic flow resulting from an acoustic pressure applied to the system. The International System of Units, SI unit of acoustic impeda ...
and its changes due to the
acoustic reflex The acoustic reflex (also known as the stapedius reflex, stapedial reflex, auditory reflex, middle-ear-muscle reflex (MEM reflex, MEMR), attenuation reflex, cochleostapedial reflex or intra-aural reflex) is an involuntary muscle contraction that ...
are dominantly determined by the structures and functional properties of the middle ear, and only marginally influenced by changes in ICP. A measurable acoustic phenomenon that originates in the inner ear would, at least in theory, allow for more precise assessment of the pressure of the peri- and endo-lymph, and consequently, of ICP.
Otoacoustic emission An otoacoustic emission (OAE) is a sound that is generated from within the inner ear. Having been predicted by Austrian astrophysicist Thomas Gold in 1948, its existence was first demonstrated experimentally by British physicist David Kemp in 19 ...
(OAE), which is a sound generated by subtle oscillations of the endo- and perilymph caused by contractions of the outer hair cells of the inner ear in response to a loud sound, seems to offer such a possibility. The sound is transmitted to the stapes, and further through the ossicles, to the tympanic membrane from which it can be detected with a sensitive microphone inserted into the ear canal. OAE is used in clinical practice to test for hearing deficits in babies and children who are too young to cooperate. The equipment can be made portable and is relatively easy to use. Two approaches have commonly utilized that increase the unfavorable signal-to-noise ratio and facilitate extraction of the OAE waveform: transient evoked otoacoustic emission (TEOAE) and distortion product otoacoustic emission (DPOAE). In a recent US patent issued to Meyerson and colleagues proposed the use of both the TEOAE and DPOAE for measurement of ICP. TEOAE is used first to determine the optimum OAE response frequency, after which the pair of pure tones is deployed in a DPOAE paradigm such that the cubic distortion product frequency equals the optimum response frequency while the ratio of frequencies f2/f1 is set to 5:4, and of intensities I2/I1 to 6:5. The inventors also proposed formulae that relate ICP to the intensity or phase of the measured OAE signal, and described how the other physiological signals or behaviors that are known to affect ICP such as small oscillations of ICP with each heartbeat, respiration, or posture changes, can be used to confirm the validity of the obtained measurements (e.g. the absence of modulation of the measured OAE phase with respiration may indicate occlusion of the cochlear aqueduct, in which case OAE cannot provide any information about ICP). There is little data up to date about the clinical utility or accuracy of otoacoustic emission as a measure of ICP. A pilot study of Frank and colleagues that evaluated different modalities of OAE in 12 healthy volunteers and 5 patients with implanted ventricular catheters for direct ICP monitoring revealed that increased ICP or conditions known to increase ICP (e.g. posture changes, abdomen compression, coughing) were associated with notable decreases (between -2.1 and -7.9SPL) in intensity of the evoked OAE. All results were however reported only as group averages, and no attempt was made to derive a quantitative one-to-one relation between the OAE intensity and ICP. This method as all other correlation-based approaches cannot be used for absolute ICP value measurement because of the impossibility of individual calibration.


Ocular measurements

Eye provides another possible window into the pressure changes in the intracranial compartment thanks to the fact that the space between the optic nerve and its sheath is a continuation of the subarachnoid space, and is consequently filled with
cerebrospinal fluid Cerebrospinal fluid (CSF) is a clear, colorless Extracellular fluid#Transcellular fluid, transcellular body fluid found within the meninges, meningeal tissue that surrounds the vertebrate brain and spinal cord, and in the ventricular system, ven ...
whose pressure is equal to intracranial pressure. Intracranial hypertension will thus manifest in an increased diameter of the optic nerve sheath and will impede the blood flow through the central retinal vein that courses within the sheath, along and in part inside of the optical nerve. The impediment of venous return causes visible changes in the eye fundus (venous engorgement, and papilledema, i.e. swelling and elevation of the optic nerve disc) that can be observed with an
ophthalmoscope Ophthalmoscopy, also called funduscopy, is a test that allows a health professional to see inside the fundus of the eye and other structures using an ophthalmoscope (or funduscope). It is done as part of an eye examination and may be done as part ...
and have therefore been used by clinicians for more than a century as signs of increased ICP. Quantitative assessment of ICP can be made noninvasively in two different ways: by measuring changes in diameter of the optic nerve sheath with an appropriate technique (ultrasound or MRI), or by using ophthalmodynamometry to determine the pressure in the central retinal vein, which is normally slightly higher (1- 2mmHg) than ICP. Intracranial hypertension also induces changes at the cellular or axonal level such as the swelling of the fibers of the optic nerve that form the innermost layer of the retina (so-called nerve fiber layer – NFL). The information provided by the classic ophthalmoscopy is however only qualitative and may be inconclusive during early phases of intracranial hypertension since it usually takes between two and four hours from the onset of ICP elevation for a papilledema to develop. A patented method that utilizes optical coherence
tomography Tomography is imaging by sections or sectioning that uses any kind of penetrating wave. The method is used in radiology, archaeology, biology, atmospheric science, geophysics, oceanography, plasma physics, materials science, cosmochemistry, ast ...
to measure the thickness of the nerve fiber layer and infers ICP from it laid claims of being able to detect the IH-induced thickening of the retina shortly after the onset of IH, but there has been no data that would support the claims or clarify the relationship between the NFL thickness and levels of ICP.


Optic nerve sheath diameter

The use of optic nerve sheath diameter (ONSD) for the assessment of ICP dates back to 1987 when Cennamo and colleagues demonstrated a linear relationship between ICP and the sheath diameter measured with a trans-orbital ultrasound probe in an A-scan mode (principally equivalent to the time-of-the-flight measurements of the cranium diameter). The original measurement method was technically difficult and unreliable because of the nearly
coaxial In geometry, coaxial means that several three-dimensional linear or planar forms share a common axis. The two-dimensional analog is ''concentric''. Common examples: A coaxial cable has a wire conductor in the centre (D), a circumferential ou ...
alignment of the optic nerve and propagation axis of the ultrasound wave, but the precision was significantly improved with the use of B-scan (or planar) ultrasound which provided longitudinal cross-section images of the optic nerve and its sheath. Since then, the method has been successfully validated in several relatively large studies that included patients with severe
head trauma A head injury is any injury that results in trauma to the skull or brain. The terms ''traumatic brain injury'' and ''head injury'' are often used interchangeably in the medical literature. Because head injuries cover such a broad scope of inj ...
,
hydrocephalus Hydrocephalus is a condition in which cerebrospinal fluid (CSF) builds up within the brain, which can cause pressure to increase in the skull. Symptoms may vary according to age. Headaches and double vision are common. Elderly adults with n ...
,
intracranial hemorrhage Intracranial hemorrhage (ICH) refers to any form of Hemorrhage, bleeding Internal bleeding, within the Human skull, skull. It can result from trauma, vascular abnormalities, hypertension, or other medical conditions. ICH is broadly categorized ...
or stroke,
liver failure Liver failure is the inability of the liver to perform its normal synthetic and metabolic functions as part of normal physiology. Two forms are recognised, acute and chronic (cirrhosis). Recently, a third form of liver failure known as acute- ...
, and climbers with
acute mountain sickness Altitude sickness, the mildest form being acute mountain sickness (AMS), is a harmful effect of high altitude, caused by rapid exposure to low amounts of oxygen at high elevation. People's bodies can respond to high altitude in different wa ...
. While the ONSD can at any given point along the optic nerve be measured with a precision of <1mm, the reliability of derived ICP levels is plagued by inter-individual variability and the dependence of ONSD magnitude on the point along the nerve at which the measurement was taken. Almost all validation studies so far have recommended that ONSD to be used for identification of patients with intracranial hypertension that requires treatment (ICP>20mmHg, i.e. ONSD>5mmHg) rather than for a measurement of ICP. ONSD measurement by ultrasound is not as precise as invasive ICP measurement but may be useful if invasive measurement is not available.


Ophthalmodynamometry or the measurement of the retinal venous outflow pressure (VOP)

Performed by applying external pressure on the sclera, for example with a
spring plunger A spring plunger or detent spring is a spring-loaded mechanical part used for indexing, positioning, and securing of objects, as well as for making objects possible to disassemble in normal use without losing parts. The spring force keeps the pin i ...
, while observing the retinal vessels through an
ophthalmoscope Ophthalmoscopy, also called funduscopy, is a test that allows a health professional to see inside the fundus of the eye and other structures using an ophthalmoscope (or funduscope). It is done as part of an eye examination and may be done as part ...
. The pressure is gradually increased until the central retinal vein begins to pulsate, which happens at the point when the applied external pressure nears the VOP and is approximately equal to ICP. The original method was described in 1925 by Baurmann and belongs to the public domain, but several modifications have been recently patented that combine the classic ophthalmodynamometry with reflectance oximetry of the
retina The retina (; or retinas) is the innermost, photosensitivity, light-sensitive layer of tissue (biology), tissue of the eye of most vertebrates and some Mollusca, molluscs. The optics of the eye create a focus (optics), focused two-dimensional ...
or ultrasound measurement of blood flow in the central retinal artery, or automate the method by adding a camera and an image processing software capable of recognizing venous pulsations from a sequence of images of the eye fundus. Evaluation in patients confirmed a strong linear relationship and clinically negligible differences (2-3mmHg) between VOP and the invasively measured ICP. Ophthalmodynamometry requires dilated pupils, a skilled physician or medic, and collaboration of the patient, which all hampers its applicability in the field. It cannot be applied in cases of ocular trauma or conditions that selectively affect the optic nerve and gives erroneously high readings in the presence of a
papilledema Papilledema or papilloedema is optic disc swelling that is caused by increased intracranial pressure due to any cause. The swelling is usually bilateral and can occur over a period of hours to weeks. Unilateral presentation is extremely rare. In ...
, which may persist long after ICP has returned to normal. ''Neurolife Non-invasive Solutions Inc.'' developed technology based on Braxton's patent. They won Purdue University's business plan competition in 2006 and were using the funds to develop iScan, its initial prototype. The approach worked on measuring ICP non-invasively by assessing changes in the retinal blood flow. However this is dependent on other factors apart from ICP, so it will be difficult to generate accuracy sufficient for clinical practice. ''Third Eye Diagnostics, Inc.'' is developing the Cerepress™, a non-invasive, hand-held intracranial pressure monitor that gathers information from the patient's un-dilated eye. The Cerepress™ measures blood pressure in the eye's central retinal vein (CRV) and blood velocity in the ophthalmic artery, which taken together highly correlate to intracranial pressure. To obtain CRV pressure, 3ED has developed a novel apparatus that simultaneously record images of the CRV and measures intraocular pressure (IOP) while the pressure in the eye is increased. A medical technician aligns the system by easily centering the field of view to patient's pupil eye. The system then contacts the patient's cornea and simultaneously collects images of the cornea and the retinal fundus. The contact force increases the IOP and momentarily compresses the CRV. At the instant of complete CRV compression, the Cerepress™ records the eye pressure, which is equivalent to CRV pressure. CRV pressure is a known to be a good correlate to ICP. This method requires an individual patient-specific calibration as every other "correlation-based" method.


See also

*
Intracranial pressure Intracranial pressure (ICP) is the pressure exerted by fluids such as cerebrospinal fluid (CSF) inside the skull and on the brain tissue. ICP is measured in millimeters of mercury ( mmHg) and at rest, is normally 7–15 mmHg for a supine adu ...
* Intracranial hypotension *
Intracranial pressure monitoring The monitoring of intracranial pressure (ICP) is used in the treatment of a number of neurological conditions ranging from severe traumatic brain injury to stroke and brain bleeds. This process is called intracranial pressure monitoring. Monitor ...
*
Headache A headache, also known as cephalalgia, is the symptom of pain in the face, head, or neck. It can occur as a migraine, tension-type headache, or cluster headache. There is an increased risk of Depression (mood), depression in those with severe ...
*
Meningitis Meningitis is acute or chronic inflammation of the protective membranes covering the brain and spinal cord, collectively called the meninges. The most common symptoms are fever, intense headache, vomiting and neck stiffness and occasion ...
*
Space adaptation syndrome Space adaptation syndrome (SAS) or space sickness is a condition experienced by as many as half of all space travelers during their adaptation to weightlessness once in orbit. It is the opposite of terrestrial motion sickness since it occurs whe ...
*
Traumatic brain injury A traumatic brain injury (TBI), also known as an intracranial injury, is an injury to the brain caused by an external force. TBI can be classified based on severity ranging from mild traumatic brain injury (mTBI/concussion) to severe traumati ...
* Cushing's triad


References

{{DEFAULTSORT:Non-invasive intracranial pressure (ICP) measurement methods Medical signs Neurotrauma Neurology procedures