HOME

TheInfoList



OR:

Microshock refers to the risk that patients undergoing medical procedures involving externally protruding intracardiac
electrical conductor In physics and electrical engineering, a conductor is an object or type of material that allows the flow of charge (electric current) in one or more directions. Materials made of metal are common electrical conductors. Electric current is gen ...
s, such as external
pacemaker An artificial cardiac pacemaker (or artificial pacemaker, so as not to be confused with the natural cardiac pacemaker) or pacemaker is a medical device that generates electrical impulses delivered by electrodes to the chambers of the heart ei ...
electrode An electrode is an electrical conductor used to make contact with a nonmetallic part of a circuit (e.g. a semiconductor, an electrolyte, a vacuum or air). Electrodes are essential parts of batteries that can consist of a variety of materials ...
s, or saline filled
catheter In medicine, a catheter (/ˈkæθətər/) is a thin tube made from medical grade materials serving a broad range of functions. Catheters are medical devices that can be inserted in the body to treat diseases or perform a surgical procedure. Cat ...
s, could suffer an electric shock causing ventricular fibrillation (VF) due to currents entering the body via these parts.


Some definitions related to micro-shock

It is important to note that microshock (or micro-shock) are not IEV defined terms and are not used in any international standard. "Micro-shock" is an otherwise imperceptible electric current applied directly, or in very close proximity, to the heart muscle of sufficient strength, frequency, and duration to cause disruption of normal cardiac function. Note: It can be safely assumed (and it usually is) that micro-shock is only possible during certain medical procedures as the electric current needs to be focused directly into the heart by some conductor inserted by invasive means for some desired medical outcome (for example Cardiac Catheterisation). Micro-shock, if it occurs, is not always lethal. “''Micro-electrocution''” is the term that should be used whenever a micro-shock causes death. “''Macro-shock''” is when a much larger current is passed through the body, usually via a skin to skin pathway, but more generally the current is not applied directly through the heart muscle. The current in macro-shock events can vary widely from being imperceptible to being extremely destructive of tissue. (see Macroshock) “''Electric Shock''” is usually referring to macro-shock. (see
Electric Shock Electrical injury is a physiological reaction caused by electric current passing through the body. The injury depends on the density of the current, tissue resistance and duration of contact. Very small currents may be imperceptible or produce a ...
) “''Electrocution''” is usually referring to a macro-shock that has caused prolonged or severe disruption of normal cardiac function - ultimately leading to death. (see
Electrocution Electrocution is death or severe injury caused by electric shock from electric current passing through the body. The word is derived from "electro" and "execution", but it is also used for accidental death. The term "electrocution" was coined ...
)


Theory

Microschock requires direct electrical connection to the heart muscle and is normally illustrated using a diagram such as Figure 1 (from TGE). (an image is to be uploaded here shortly) In this scenario the patient has inadvertently contacted both a source of current (it does not have to be AC, as shown) as well as a common return pathway during an invasive cardiac medical procedure. If the current flowing is below the threshold of perception, or the patient is sedated, or anaesthetized, there may be no pain or reflex response of either arm. If the current flow continues for sufficient time, at sufficient strength, the patient may die. Because of the low current and lack of patient response, this death may be unexpected, and without any obvious cause. In practice, however, this has never been proven to have happened. To a novice, however, this scenario looks incredibly dangerous, and it is therefore worth examining in some detail. Firstly, let’s follow the current path. There is a generic source of current. This source can be either large or small, as only a small voltage is required to drive the low current for micro-shock. Such sources might be, a wall socket, a faulty item of equipment, an inappropriate item of equipment, a poorly designed item of equipment, or an item of equipment designed to deliver current into the body. Our patient has unfortunately contacted one such source and current is dispersing through their right arm and upper torso, to eventually converge on a catheter (as labelled – but it could be a lead or wire) that is placed into their heart. This concentration of current flow at the heart muscle is the danger from micro-shock. If the catheter is conductive and insulated, the current may follow the catheter, emerging through the skin into some other item of equipment. For the circuit as shown to be complete, the conductive part of the catheter needs to also be connected to ground within this equipment. Finally, the hazardous circuit is complete – current can flow and if it continues the patient is in mortal danger. Again, while this is theoretically possible this has never been proven to have actually happened. So, why is this situation not arisen? The details are explored below but the basic presumptions that a patient will simultaneously contact the mains, with a catheter inserted, using equipment that grounds the catheter, is very, very unlikely. Also most electrical connections made in or around a patients heart will be those of medical electrical equipment. Medical electrical equipment, which have such applied parts, are constructed to strict standards that limit the allowable currents flowing via such connections (applied parts). This ensures the safety of the patient.


History

There has never been a documented case of microshock. A U.S. Senate inquiry in the early 1970s, sparked by exaggerated reports of thousands of U.S hospital patients dying of microshock, heard expert testimony about the effect. A review of the evidence in the early 2000s found that not a single case had been reported in the 30 years since the Senate inquiry.Dyro Regular checks of the FDA's MAUDE database also show no evidence of this risk being manifest, before or since the review. Based on studies with dogs by Prof Leslie Geddes in the middle of last century, it is theorised that a current as low as 10 μA (
microampere The ampere (, ; symbol: A), often shortened to amp,SI supports only the use of symbols and deprecates the use of abbreviations for units. is the unit of electric current in the International System of Units (SI). One ampere is equal to elect ...
) directly through the
heart The heart is a muscular organ found in most animals. This organ pumps blood through the blood vessels of the circulatory system. The pumped blood carries oxygen and nutrients to the body, while carrying metabolic waste such as carbon diox ...
, may send a human patient directly into
ventricular fibrillation Ventricular fibrillation (V-fib or VF) is an abnormal heart rhythm in which the Ventricle (heart), ventricles of the heart Fibrillation, quiver. It is due to disorganized electrical conduction system of the heart, electrical activity. Ventricular ...
. Of course, the exact outcome is dependent on the duration of the current, the exact position of contact, the frequency of current oscillation, and the timing of the shock with the hearts rhythm e.g R on T phenomenon. It is feared that such a small current may be introduced unwittingly, and unobserved, creating a very perilous situation for the patient. To guard against this slim theoretical possibility then, modern medical devices include a range of protective measures to limit current in cardiac-connected circuits to the assumed safe levels of below 10 μA (
microampere The ampere (, ; symbol: A), often shortened to amp,SI supports only the use of symbols and deprecates the use of abbreviations for units. is the unit of electric current in the International System of Units (SI). One ampere is equal to elect ...
) . These measures include isolated patient connections, high impedance connections and current limiting circuits. Despite the in-built protections, and lack of observed incidents, microshock continues to be a concern to many practitioners of the fields of Biomedical and Clinical Engineering. Despite the evidence of decades of absence of reports, in any condition where electrical conductors are run into the body in proximity of the heart (i.e. cardiac catheterizations) precautions are still taken to ensure hazardous current is not introduced through these conductors and it is still regarded as a high risk activity.


See also

* Macroshock *
Electric Shock Electrical injury is a physiological reaction caused by electric current passing through the body. The injury depends on the density of the current, tissue resistance and duration of contact. Very small currents may be imperceptible or produce a ...


Notes


References

*Gross J (2005) Less Jolts from Your Volts: Electrical Safety in the Operating Room. ASA Refresher Courses in Anesthesiology. 33(1):101-114 *Ridgway M The Great Debate on Electrical Safety - in Retrospect in Clinical Engineering Handbook, 1st Edition, Chapter 65, J Dyro (ed) Academic Press 2004 *O'Meley P L Who's Afraid of Microshock - presentation to SMBE NSW Conference, Albury NSW, 2011 *Hsu J The Hypertextbook http://hypertextbook.com/facts/2000/JackHsu.shtml accessed 23 July 2013 *FDA MAUDE Database http://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfmaude/TextSearch.cfm accessed 25 July 2013 *Road Safety Report http://roadsafety.transport.nsw.gov.au/downloads/fatality_rate_1908_to_2009.pdf{{dead link, date=January 2018 , bot=InternetArchiveBot , fix-attempted=yes accessed 25 July 2013 *IEV Electropedia produced by the International Electrotechnical Commission (IEC) http://www.electropedia.org *IEC/TS 60479-1 - Effects of current on human beings and livestock https://webstore.iec.ch/publication/25402 Biomedical engineering Cardiac electrophysiology