A mass spectrum is a
histogram plot of intensity vs. ''
mass-to-charge ratio'' (''m/z'') in a chemical sample, usually acquired using an instrument called a ''
mass spectrometer
Mass spectrometry (MS) is an analytical technique that is used to measure the mass-to-charge ratio of ions. The results are presented as a '' mass spectrum'', a plot of intensity as a function of the mass-to-charge ratio. Mass spectrometry is u ...
''. Not all mass spectra of a given substance are the same; for example, some mass spectrometers break the analyte molecules into ''
fragments''; others observe the intact molecular masses with little fragmentation. A mass spectrum can represent many different types of information based on the type of mass spectrometer and the specific experiment applied. Common fragmentation processes for organic molecules are the ''
McLafferty rearrangement'' and ''
alpha cleavage''. Straight chain alkanes and alkyl groups produce a typical series of peaks: 29 (CH
3CH
2+), 43 (CH
3CH
2CH
2+), 57 (CH
3CH
2CH
2CH
2+), 71 (CH
3CH
2CH
2CH
2CH
2+) etc.
X-axis: ''m/z'' (mass-to-charge ratio)
The
x-axis of a mass spectrum represents a relationship between the mass of a given ion and the number of elementary charges that it carries. This is written as the
IUPAC
The International Union of Pure and Applied Chemistry (IUPAC ) is an international federation of National Adhering Organizations working for the advancement of the chemical sciences, especially by developing nomenclature and terminology. It is ...
standard ''m/z'' to denote the quantity formed by dividing the mass of an ion by the
unified atomic mass unit
The dalton or unified atomic mass unit (symbols: Da or u) is a non-SI unit of mass widely used in physics and chemistry. It is defined as of the mass of an unbound neutral atom of carbon-12 in its nuclear and electronic ground state and at r ...
and by its charge number (positive absolute value).
This has been referred to as a
mass-to-charge ratio, although in some ways it does not fit this description. The IUPAC Gold Book gives an example:
"''for the ion C
7H
72+, m/z equals 45.5''".
Since a mass spectrum ''x''-axis represents a relationship between the ion mass and the number of elementary charges that a given ion carries it contains mass information that may be extracted by a mass spectrometrist.
Alternative x-axis notations
There are several alternatives to the standard ''m/z'' notation that appear in the literature; however, these are not currently accepted by standards organizations and most journals. ''m/e'' appears in older historical literature. A label more consistent with the
IUPAC green book and
ISO 31
ISO 31 ( Quantities and units, International Organization for Standardization, 1992) is a superseded international standard concerning physical quantities, units of measurement, their interrrelationships and their presentation. It was revised a ...
conventions is ''m/Q'' or ''m/q'' where ''m'' is the symbol for mass and ''Q'' or ''q'' the symbol for charge with the units u/e or Da/e. This notation is not uncommon in the physics of mass spectrometry but is rarely used as the abscissa of a mass spectrum. It was also suggested to introduce a new unit
thomson (Th) as a unit of ''m/z'', where 1 Th = 1 u/e. According to this convention, mass spectra x axis could be labeled ''m/z'' (Th) and negative ions would have negative values. This notation is rare and not accepted by
IUPAC
The International Union of Pure and Applied Chemistry (IUPAC ) is an international federation of National Adhering Organizations working for the advancement of the chemical sciences, especially by developing nomenclature and terminology. It is ...
or any other standards organisation.
History of x-axis notation
In 1897 the mass-to-charge ratio
of the
electron
The electron (, or in nuclear reactions) is a subatomic particle with a negative one elementary electric charge. Electrons belong to the first generation of the lepton particle family,
and are generally thought to be elementary partic ...
was first measured by
J. J. Thomson. By doing this he showed that the electron, which was postulated before in order to explain electricity, was in fact a particle with a mass and a charge and that its mass-to-charge ratio was much smaller than the one for the hydrogen ion H
+. In 1913 he measured the mass-to-charge ratio of
ions with an instrument he called a parabola spectrograph. Although this data was not represented as a modern mass spectrum, it was similar in meaning. Eventually there was a change to the notation as ''m/e'' giving way to the current standard of ''m/z''.
Early in mass spectrometry research the
resolution of mass spectrometers did not allow for accurate mass determination.
Francis William Aston won the Nobel prize in Chemistry in 1922. "For his discovery, by means of his mass spectrograph, of isotopes, in a large number of non-radioactive elements, and for his enunciation of the
Whole Number Rule." In which he stated that all atoms (including isotopes) follow a whole-number rule This implied that the masses of atoms were not on a scale but could be expressed as integers (in fact multiple charged ions were rare, so for the most part the ratio was whole as well). There have been several suggestions (e.g. the unit thomson) to change the official mass spectrometry nomenclature
to be more internally consistent.
Y-axis: signal intensity
The
''y''-axis of a mass spectrum represents signal intensity of the ions. When using counting detectors the intensity is often measured in counts per second (cps). When using analog detection electronics the intensity is typically measured in volts. In
FTICR
Fourier-transform ion cyclotron resonance mass spectrometry is a type of mass analyzer (or mass spectrometer) for determining the mass-to-charge ratio (''m''/''z'') of ions based on the cyclotron frequency of the ions in a fixed magnetic field. T ...
and
Orbitraps the
frequency domain
In physics, electronics, control systems engineering, and statistics, the frequency domain refers to the analysis of mathematical functions or signals with respect to frequency, rather than time. Put simply, a time-domain graph shows how a ...
signal (the ''y''-axis) is related to the
power (~amplitude squared) of the signal
sine wave (often reduced to an
rms power); however, the axis is usually not labeled as such for many reasons. In most forms of mass spectrometry, the intensity of ion current measured by the spectrometer does not accurately represent relative abundance, but correlates loosely with it. Therefore, it is common to label the ''y''-axis with "arbitrary units".
Y-axis and relative abundance
Signal intensity may be dependent on many factors, especially the nature of the molecules being analyzed and how they ionize. The efficiency of ionization varies from molecule to molecule and from ion source to ion source. For example, in electrospray sources in positive ion mode a quaternary amine will ionize exceptionally well whereas a large hydrophobic alcohol will most likely not be seen no matter how concentrated. In an EI source these molecules will behave very differently. Additionally there may be factors that affect ion transmission disproportionally between ionization and detection.
On the detection side there are many factors that can also affect signal intensity in a non-proportional way. The size of the ion will affect the velocity of impact and with certain detectors the velocity is proportional to the signal output. In other detection systems, such as
FTICR
Fourier-transform ion cyclotron resonance mass spectrometry is a type of mass analyzer (or mass spectrometer) for determining the mass-to-charge ratio (''m''/''z'') of ions based on the cyclotron frequency of the ions in a fixed magnetic field. T ...
, the number of charges on the ion are more important to signal intensity. In
Fourier transform ion cyclotron resonance and
Orbitrap type mass spectrometers the signal intensity (Y-axis) is related to the amplitude of the
free induction decay signal. This is fundamentally a power relationship (amplitude squared) but often computed as an
ms For decaying signals the rms is not equal to the average amplitude. Additionally the damping constant (decay rate of the signal in the fid) is not the same for all ions. In order to make conclusions about relative intensity a great deal of knowledge and care is required.
A common way to get more quantitative information out of a mass spectrum is to create a standard curve to compare the sample to. This requires knowing what is to be quantitated ahead of time, having a standard available and designing the experiment specifically for this purpose. A more advanced variation on this is the use of an
internal standard which behaves very similarly to the analyte. This is often an isotopically labeled version of the analyte. There are forms of mass spectrometry, such as
accelerator mass spectrometry
Accelerator mass spectrometry (AMS) is a form of mass spectrometry that accelerates ions to extraordinarily high kinetic energies before mass analysis. The special strength of AMS among the mass spectrometric methods is its power to separate a r ...
that are designed from the bottom up to be quantitative.
Spectral skewing
Spectral skewing is the change in relative intensity of mass spectral peaks due to the changes in concentration of the
analyte
An analyte, component (in clinical chemistry), or chemical species is a substance or chemical constituent that is of interest in an analytical procedure. The purest substances are referred to as analytes, such as 24 karat gold, NaCl, water, et ...
in the
ion source as the mass spectrum is scanned. This situation occurs routinely as
chromatographic components
elute
In analytical and organic chemistry, elution is the process of extracting one material from another by washing with a solvent; as in washing of loaded ion-exchange resins to remove captured ions.
In a liquid chromatography experiment, for exam ...
into a continuous ion source.
[Watson, J. THrock, Sparkman,O David.Introduction to Mass Spectrometry.John Wiley & Sons, Inc. 4th Edition,2007. Page:113] Spectral skewing is not observed in
ion trap
An ion trap is a combination of electric and/or magnetic fields used to capture charged particles — known as ions — often in a system isolated from an external environment. Atomic and molecular ion traps have a number of applications in ph ...
(
quadrupole
A quadrupole or quadrapole is one of a sequence of configurations of things like electric charge or current, or gravitational mass that can exist in ideal form, but it is usually just part of a multipole expansion of a more complex structure ref ...
(this has been seen also in
QMS) or magnetic) or
time-of-flight
Time of flight (ToF) is the measurement of the time taken by an object, particle or wave (be it acoustic, electromagnetic, etc.) to travel a distance through a medium. This information can then be used to measure velocity or path length, or as a w ...
(TOF)
mass analyzers because potentially all
ions formed in operational cycle (a snapshot in time) of the instrument are available for detection.
See also
*
Kendrick mass
References
External links
* Quantities, Units and Symbols in Physical Chemistry
IUPAC green book
An introductory video on Mass SpectrometryThe Royal Society of Chemistry
*
NIST
The National Institute of Standards and Technology (NIST) is an agency of the United States Department of Commerce whose mission is to promote American innovation and industrial competitiveness. NIST's activities are organized into physical sc ...
br>
Standard Reference Database 1A v17
{{DEFAULTSORT:Mass Spectrum
Mass spectrometry