Automatic Exposure Control (AEC) is an X-ray exposure termination device. A medical radiographic exposure is always initiated by a human operator but an AEC detector system may be used to terminate the exposure when a predetermined amount of radiation has been received. The intention of AEC is to provide consistent x-ray image exposure, whether to film, a digital detector or a
CT scan A computed tomography scan (CT scan; formerly called computed axial tomography scan or CAT scan) is a medical imaging technique used to obtain detailed internal images of the body. The personnel that perform CT scans are called radiographers ...
ner. AEC systems may also automatically set exposure factors such as the X-ray tube current and voltage in a CT.


Operation


Projectional Radiography

In projectional radiography an AEC system uses one or more physically thin radiation ionization chambers (the "AEC detector") which is positioned between the X-ray source and the x-ray receptor. Where low energy x-rays are used such as in
mammography Mammography (also called mastography) is the process of using low-energy X-rays (usually around 30 kVp) to examine the human breast for diagnosis and screening. The goal of mammography is the early detection of breast cancer, typically through d ...
the AEC detector is placed behind the image receptor to avoid creating a shadow. In early radiographic AEC systems, a large paddle (17" x 17") of transparent lucite was sandwiched between rare earth screens, which emitted photons when excited by X-rays. The individual lucite sections were open on one end, and a solenoid was used to select one of three, or a combination of shutters that allowed the generated light into a
Photomultiplier tube Photomultiplier tubes (photomultipliers or PMTs for short) are extremely sensitive detectors of light in the ultraviolet, visible, and near-infrared ranges of the electromagnetic spectrum. They are members of the class of vacuum tubes, more specif ...
. The output of the PMT was then converted to a signal, which ramped in the positive direction until a preprogrammed threshold was reached. At this point, the X-ray generator terminated the exposure. This method is no longer used, and has been replaced by the iontomat. In an iontomat, a weak ionization signal resulting from the radiographic X-rays passing through it are integrated as a ramp shaped voltage waveform. This ramp signal rises until it matches a pre-set threshold. At this point the x-ray exposure is terminated by the X-ray generator. AEC devices are calibrated to ensure that similar exams have linearity in
optical density Absorbance is defined as "the logarithm of the ratio of incident to transmitted radiant power through a sample (excluding the effects on cell walls)". Alternatively, for samples which scatter light, absorbance may be defined as "the negative lo ...
.


Computed Tomography

Modern computed tomography (CT) scanners have AEC systems which aim to maintain image quality for patients of varying sizes, whilst keeping doses
as low as reasonably practicable ALARP ("as low as reasonably practicable"), or ALARA ("as low as reasonably achievable"), is a principle in the regulation and management of safety-critical and safety-involved systems. The principle is that the residual risk shall be reduced as f ...
. The systems are also designed to maintain quality with the varying size and attenuation of an individual patient over their length. Implementations vary between manufacturers, some systems are based on a desired noise level in the image, while others are based on a specified reference output (milliampere second,
mAs Mas, Más or MAS may refer to: Film and TV * Más y Menos, fictional superhero characters, from the Teen Titans animated television series * Más (Breaking Bad), "Más" (''Breaking Bad''), a season three episode of ''Breaking Bad'' Songs * Más ( ...
). CT AEC systems use the initial "scanogram", a fixed angle planning view, to determine the relative size of the patient, and variation over their length. The
tube Tube or tubes may refer to: * ''Tube'' (2003 film), a 2003 Korean film * ''The Tube'' (TV series), a music related TV series by Channel 4 in the United Kingdom * "Tubes" (Peter Dale), performer on the Soccer AM television show * Tube (band), a ...
output is then adjusted for overall size. The output is also typically modulated for each rotation in response to changes in attenuation over patient length. Some systems adjust output during each rotation, which is known as rotational modulation, based on measured attenuation in the previous rotation.


Advantages

Because patients vary in size and shape, an AEC device is very useful in achieving consistent x-ray film densities, which can be difficult when manually setting exposure factors without AEC.


Disadvantages

AEC devices are susceptible to operator error (usually due to mispositioned anatomy or having the incorrect AEC chamber selected). Prosthetic devices such as total hip hardware can also cause the selected ionization chamber to overexpose the image receptor. This is due to the absorption of the x-ray beam into the metal of the hardware as opposed to exposing the ionization chamber.