HOME

TheInfoList



OR:

In a
chemical analysis Analytical chemistry studies and uses instruments and methods to separate, identify, and quantify matter. In practice, separation, identification or quantification may constitute the entire analysis or be combined with another method. Separa ...
, the internal standard method involves adding the same amount of a
chemical substance A chemical substance is a unique form of matter with constant chemical composition and characteristic properties. Chemical substances may take the form of a single element or chemical compounds. If two or more chemical substances can be com ...
to each sample and
calibration In measurement technology and metrology, calibration is the comparison of measurement values delivered by a device under test with those of a calibration standard of known accuracy. Such a standard could be another measurement device of known ...
solution. The internal standard responds proportionally to changes in the
analyte An analyte, component (in clinical chemistry), titrand (in titrations), or chemical species is a substance or chemical constituent that is of interest in an analytical procedure. The remainder of the sample is called the matrix. The procedure ...
and provides a similar, but not identical, measurement signal. It must also be absent from the sample matrix to ensure there is no other source of the internal standard present. Taking the ratio of analyte signal to internal standard signal and plotting it against the analyte concentrations in the calibration solutions will result in a
calibration curve In analytical chemistry, a calibration curve, also known as a standard curve, is a general method for determining the concentration of a substance in an unknown sample by comparing the unknown to a set of standard samples of known concentration. ...
. The calibration curve can then be used to calculate the analyte concentration in an unknown sample. Selecting an appropriate internal standard accounts for random and systematic sources of uncertainty that arise during sample preparation or instrument fluctuation. This is because the ratio of analyte relative to the amount of internal standard is independent of these variations. If the measured value of the analyte is erroneously shifted above or below the actual value, the internal standard measurements should shift in the same direction. Ratio plot provides good way of compensation of detector sensitivity variation, but may be biased and should be replaced by Relative concentration/Relative calibration calculations if the reason of response variability is in different mass of analysed sample and traditional (not internal standard)
calibration curve In analytical chemistry, a calibration curve, also known as a standard curve, is a general method for determining the concentration of a substance in an unknown sample by comparing the unknown to a set of standard samples of known concentration. ...
of any
analyte An analyte, component (in clinical chemistry), titrand (in titrations), or chemical species is a substance or chemical constituent that is of interest in an analytical procedure. The remainder of the sample is called the matrix. The procedure ...
is not linear through origin.


History

The earliest recorded use of the internal standard method dates back to Gouy's flame spectroscopy work in 1877, where he used an internal standard to determine if the excitation in his flame was consistent. His experimental procedure was later reintroduced in the 1940s, when recording flame photometers became readily available. The use of internal standards continued to grow, being applied to a wide range of analytical techniques including nuclear magnetic resonance (NMR) spectroscopy,
chromatography In chemical analysis, chromatography is a laboratory technique for the Separation process, separation of a mixture into its components. The mixture is dissolved in a fluid solvent (gas or liquid) called the ''mobile phase'', which carries it ...
, and inductively coupled plasma
spectroscopy Spectroscopy is the field of study that measures and interprets electromagnetic spectra. In narrower contexts, spectroscopy is the precise study of color as generalized from visible light to all bands of the electromagnetic spectrum. Spectro ...
.


Applications


Nuclear magnetic resonance spectroscopy

In NMR spectroscopy, e.g. of the nuclei 1H, 13C and 29Si, frequencies depend on the magnetic field, which is not the same across all experiments. Therefore, frequencies are reported as relative differences to
tetramethylsilane Tetramethylsilane (abbreviated as TMS) is the organosilicon compound with the formula Si(CH3)4. It is the simplest tetraorganosilane. Like all silanes, the TMS framework is tetrahedral. TMS is a building block in organometallic chemistry but als ...
(TMS), an internal standard that George Tiers proposed in 1958 and that the
International Union of Pure and Applied Chemistry The International Union of Pure and Applied Chemistry (IUPAC ) is an international federation of National Adhering Organizations working for the advancement of the chemical sciences, especially by developing nomenclature and terminology. It is ...
has since endorsed. The relative difference to TMS is called
chemical shift In nuclear magnetic resonance (NMR) spectroscopy, the chemical shift is the resonant frequency of an atomic nucleus relative to a standard in a magnetic field. Often the position and number of chemical shifts are diagnostic of the structure of ...
. TMS works as an ideal standard because it is relatively inert and its identical methyl protons produce a strong upfield signal, isolated from most other protons. It is soluble in most organic solvents and is removable via distillation due to its low boiling point. In practice, the difference between the signals of common solvents and TMS are known. Therefore, no TMS needs to be added to commercial deuterated solvents, as modern instruments are capable of detecting the small quantities of protonated solvent present. By specifying the lock solvent to be used, modern spectrometers are able to correctly reference the sample; in effect, the solvent itself serves as the internal standard.


Chromatography

In chromatography, internal standards are used to determine the concentration of other analytes by calculating response factor. The selected internal standard should have a similar retention time and
derivatization Derivatization is a technique used in chemistry which converts a chemical compound into a product (the reaction's derivate) of similar chemical structure, called a derivative. Generally, a specific functional group of the compound participates ...
. It must be stable and not interfere with the sample components. This mitigates the uncertainty that can occur in preparatory steps such as sample injection. In gas chromatography-mass spectrometry (GC-MS), deuterated compounds with similar structures to the analyte commonly act as effective internal standards. However, there are non-deuterated internal standards such as
norleucine Norleucine (abbreviated as Nle) is an amino acid with the formula CH3(CH2)3CH(NH2)CO2H. A systematic name for this compound is 2-aminohexanoic acid. The compound is an isomer of the more common amino acid leucine. Like most other α-amino acids, ...
, which is popular in the analysis of amino acids because it can be separated from accompanying peaks. Selecting an internal standard for liquid chromatography-mass spectrometry (LC-MS) depends on the employed ionization method. The internal standard needs a comparable ionization response and fragmentation pattern to the analyte. LC-MS internal standards are often isotopically analogous to the structure of the analyte, using
isotope Isotopes are distinct nuclear species (or ''nuclides'') of the same chemical element. They have the same atomic number (number of protons in their Atomic nucleus, nuclei) and position in the periodic table (and hence belong to the same chemica ...
s such as deuterium (2H), 13C, 15N and 18O.


Inductively coupled plasma

Selecting an internal standard in inductively coupled plasma spectroscopy can be difficult, because signals from the sample matrix can overlap with those belonging to the analyte.
Yttrium Yttrium is a chemical element; it has Symbol (chemistry), symbol Y and atomic number 39. It is a silvery-metallic transition metal chemically similar to the lanthanides and has often been classified as a "rare-earth element". Yttrium is almost a ...
is a common internal standard that is naturally absent in most samples. It has both a mid-range mass and emission lines that don't interfere with many analytes. The intensity of the yttrium signal is what the signal from the analyte gets compared to. In Inductively coupled plasma-mass spectrometry (ICP-MS), species with a similar mass to the analyte usually serve as good internal standards, though not in every case. Factors that also contribute to the effectiveness of an internal standard in ICP-MS include how close its
ionization potential In physics and chemistry, ionization energy (IE) is the minimum energy required to remove the most loosely bound electron of an isolated gaseous atom, positive ion, or molecule. The first ionization energy is quantitatively expressed as :X(g) ...
, change in
enthalpy Enthalpy () is the sum of a thermodynamic system's internal energy and the product of its pressure and volume. It is a state function in thermodynamics used in many measurements in chemical, biological, and physical systems at a constant extern ...
, and change in
entropy Entropy is a scientific concept, most commonly associated with states of disorder, randomness, or uncertainty. The term and the concept are used in diverse fields, from classical thermodynamics, where it was first recognized, to the micros ...
are to the analyte. Inductively coupled plasma-optical emission spectroscopy (ICP-OES) internal standards can be selected by observing how the analyte and internal standard signals change with varying experimental conditions. This includes making adjustments to the sample matrix or instrumentation settings and evaluating whether the selected internal standard is reacting in the same way the analyte is.


Example of internal standard method

One way to visualize the internal standard method is to create one calibration curve that doesn't use the method and one calibration curve that does. Suppose there are known concentrations of nickel in a set of calibration solutions: 0 ppm, 1.6 ppm, 3.2 ppm, 4.8 ppm, 6.4 ppm, and 8 ppm. Each solution also has 5 ppm yttrium to act as an internal standard. If these solutions are measured using ICP-OES, the intensity of the yttrium signal should be consistent across all solutions. If not, the intensity of the nickel signal is likely imprecise as well. The calibration curve that does not use the internal standard method ignores the uncertainty between measurements. The
coefficient of determination In statistics, the coefficient of determination, denoted ''R''2 or ''r''2 and pronounced "R squared", is the proportion of the variation in the dependent variable that is predictable from the independent variable(s). It is a statistic used in t ...
(R2) for this plot is 0.9985. In the calibration curve that uses the internal standard, the y-axis is the ratio of the nickel signal to the yttrium signal. This ratio is unaffected by uncertainty in the nickel measurements, as it should affect the yttrium measurements in the same way. This results in a higher R2, 0.9993.


References

{{Analytical chemistry Analytical chemistry