Soft Ionization by Chemical Reaction in Transfer is a method for ionizing small
organic compound Some chemical authorities define an organic compound as a chemical compound that contains a carbon–hydrogen or carbon–carbon bond; others consider an organic compound to be any chemical compound that contains carbon. For example, carbon-co ...
s at ambient
atmospheric pressure Atmospheric pressure, also known as air pressure or barometric pressure (after the barometer), is the pressure within the atmosphere of Earth. The standard atmosphere (symbol: atm) is a unit of pressure defined as , which is equivalent to 1,013. ...
. It is used for ion generation in mass spectrometers (MS). The technique is often abbreviated as SICRIT, a trademark of Plasmion GmbH, which commercialized the technique in the 2020s.


Ionization technique

Sicrit ionizes molecules through a technique known as
dielectric barrier discharge Dielectric-barrier discharge (DBD) is the electrical discharge between two electrodes separated by an insulating dielectric barrier. Originally called silent (inaudible) discharge and also known as ozone production discharge or partial disc ...
. This process involves creating an
electrical discharge In electromagnetism, an electric discharge is the release and transmission of electricity in an applied electric field through a medium such as a gas (i.e., an outgoing flow of electric current through a non-metal medium).American Geophysical U ...
between two
electrode An electrode is an electrical conductor used to make contact with a nonmetallic part of a circuit (e.g. a semiconductor, an electrolyte, a vacuum or a gas). In electrochemical cells, electrodes are essential parts that can consist of a varie ...
s that are separated by a
dielectric In electromagnetism, a dielectric (or dielectric medium) is an Insulator (electricity), electrical insulator that can be Polarisability, polarised by an applied electric field. When a dielectric material is placed in an electric field, electric ...
. The electrodes are a stainless-steel tube placed inside a copper tube, with a thin
quartz Quartz is a hard, crystalline mineral composed of silica (silicon dioxide). The Atom, atoms are linked in a continuous framework of SiO4 silicon–oxygen Tetrahedral molecular geometry, tetrahedra, with each oxygen being shared between two tet ...
tube separating them. The sample containing the molecules to be ionized flows through the quartz tube separating the electrodes. A
voltage Voltage, also known as (electrical) potential difference, electric pressure, or electric tension, is the difference in electric potential between two points. In a Electrostatics, static electric field, it corresponds to the Work (electrical), ...
is applied to the electrodes; the resulting electrical discharge ignites a
cold plasma Cold is the presence of low temperature, especially in the atmosphere. In common usage, cold is often a subjective perception. A lower bound to temperature is absolute zero, defined as 0.00K on the Kelvin scale, an absolute thermodynamic ...
which ionizes the sample. Unlike other ambient ionization methods such as DART, DAPPI or
DESI Desi ( or or ; Hindustani language, Hindustani: देसी , , ) also Deshi, is a loose term used to describe the ethnic groups in South Asia, peoples, culture of South Asia, cultures, and products of the Indian subcontinent and their Sout ...
, SICRIT takes place directly in line with the inlet system of the MS and thus in a continuous flow:Several reaction pathways have been identified that lead to ionization during this process: * The proton transfer via hydronium clusters, similar to spray-based methods (c. ESI). * The hydronium cluster mechanism as with chemical methods (cf. APCI). * The
radical Radical (from Latin: ', root) may refer to: Politics and ideology Politics *Classical radicalism, the Radical Movement that began in late 18th century Britain and spread to continental Europe and Latin America in the 19th century *Radical politics ...
cation pathway: After an initial electron ionization of gas, solvent, and analyte, the formed radical cations are stabilized by abstraction of hydrogen atoms from other reactants, such as solvents, resulting mainly in a protonated molecule. * A reaction pathway where charge transfer occurs through reactive species and UV radiation (
photoionization Photoionization is the physical process in which an ion is formed from the interaction of a photon with an atom or molecule. Cross section Not every interaction between a photon and an atom, or molecule, will result in photoionization. The prob ...
). Regardless of the specific reaction pathway, almost exclusively protonated species +H are generated during ionization. Since the analytes do not come into direct contact with the plasma during ionization, but rather charge transfer occurs via reactive species and UV radiation, the molecules remain intact and fragmentation is avoided. Consequently, SICRIT is a very "soft" ionization method.


Characteristics

The SICRIT technology decouples sample delivery from the ionization process. Through the flow principle, the sample is directly drawn into the high vacuum behind the MS inlet and ionized on its way into the inlet. The electrode geometry is chosen so that, under the given physical parameters (pressure, ignition voltage, gas constant, see
Paschen's law Paschen's law is an equation that gives the breakdown voltage, that is, the voltage necessary to start a discharge or electric arc, between two electrodes in a gas as a function of pressure and gap length. It is named after Friedrich Paschen who ...
), flexibility regarding the plasma medium is ensured. This allows the generation of stable cold plasma even in ambient air. In the simplest application, ambient air can be directly analyzed. Ionization in flow enables high ion transmission since ion loss is significantly reduced compared to the most commonly used spraying methods in mass spectrometry (cf. ESI, APCI). Consequently, this leads to an increase in sensitivity compared to these methods. Ionization in flow also allows for real-time measurement without the need for sample preparation. Simple screening applications, especially for
volatile organic compound Volatile organic compounds (VOCs) are organic compounds that have a high vapor pressure at room temperature. They are common and exist in a variety of settings and products, not limited to Indoor mold, house mold, Upholstery, upholstered furnitur ...
analysis, can be easily implemented since the usual sample preparation (including crushing, extraction, purification, etc.) becomes obsolete for simple screening applications. The different reaction pathways in plasma ionization broaden the spectrum of ionizable substances. This means that compared to ionization methods that only allow for single reaction pathways, a wider polarity range of analytes is covered, and nonpolar substances such as hexane can be ionized. The low fragmentation ionization allows for identification based on the molecular mass as protonated +H species. This can be particularly useful in combination with high-resolution mass spectrometers such as time-of-flight mass spectrometers (TOF-MS) or Orbitrap-MS for non-target analysis, where the entire substance spectrum of a sample is captured based on the exact mass of the molecules. Thus, SICRIT can be used in combination with both
liquid chromatography In chemical analysis, chromatography is a laboratory technique for the separation of a mixture into its components. The mixture is dissolved in a fluid solvent (gas or liquid) called the ''mobile phase'', which carries it through a system ...
(LC) as well as
gas chromatography Gas chromatography (GC) is a common type of chromatography used in analytical chemistry for Separation process, separating and analyzing compounds that can be vaporized without Chemical decomposition, decomposition. Typical uses of GC include t ...
(GC). This enables the performance of both LC-MS and GC-MS analyses on the same mass spectrometer and the establishment of a unified database for comparing data from these otherwise instrumentally separated separation and detection methods.


Instruments and application


Direct screening

As ambient ionization, SICRIT technology enables direct, real-time gas phase measurement using a mass spectrometer. The sample is positioned directly in front of the SICRIT source without any preparation. One application area is the measurement of aromatic compounds.


Chromatography couplings

The SICRIT ion source allows for coupling with various types of chromatography (GC,
HPLC High-performance liquid chromatography (HPLC), formerly referred to as high-pressure liquid chromatography, is a technique in analytical chemistry used to separate, identify, and quantify specific components in mixtures. The mixtures can origina ...
, SFC, etc.) as interface technology to any atmospheric pressure mass spectrometer (LC-MS). The ionization with its characteristics (see above) is not influenced by the coupling, allowing the same ionization method to be used for different chromatography couplings. The ability to couple gas chromatography with a low-fragmentation ionization technique on an LC-MS, for example, can be utilized in the analysis of saturated
hydrocarbon In organic chemistry, a hydrocarbon is an organic compound consisting entirely of hydrogen and carbon. Hydrocarbons are examples of group 14 hydrides. Hydrocarbons are generally colourless and Hydrophobe, hydrophobic; their odor is usually fain ...
s. Electron impact ionization commonly used in GC-MS leads to difficult-to-interpret fragmentation spectra, while plasma ionization provides fragmentation-free spectra. Thus, the DBD plasma with its broad ionization range opens up new fields of application possibilities for LC mass spectrometers in residue analysis, such as pesticides, where gas chromatographic separation is the method of choice and plasma ionization achieves very low detection limits.


Chemical imaging

In combination with appropriate sample preparation and instrumentation, the SICRIT ion source can also be used for imaging mass spectrometry. The standard procedure typically involves elaborate sample preparation combined with laser desorption/ionization (e.g.,
MALDI In mass spectrometry, matrix-assisted laser desorption/ionization (MALDI) is an ionization technique that uses a laser energy-absorbing matrix to create ions from large molecules with minimal fragmentation. It has been applied to the analysis of ...
or ''atmospheric pressure MALDI''), allowing spatial visualization of biomolecules, for example, in tissue sections.Efstathios A. Elia, Marcel Niehaus, Rory T. Steven, Jan-Christoph Wolf, Josephine Bunch: ''Atmospheric Pressure MALDI Mass Spectrometry Imaging Using In-Line Plasma Induced Postionization''. In: ''Analytical Chemistry''. Volume 92, No. 23, 2020, ISSN 0003-2700, p. 15285–15290, doi:10.1021/acs.analchem.0c03524. The use of the SICRIT source for additional in-line post-ionization in AP-MALDI experiments can result in significant signal enhancement in the detection of
metabolite In biochemistry, a metabolite is an intermediate or end product of metabolism. The term is usually used for small molecules. Metabolites have various functions, including fuel, structure, signaling, stimulatory and inhibitory effects on enzymes, c ...
s in biological sample material or enables detection of small (bio)molecules that are not addressable using MALDI alone. Furthermore, the SICRIT ion source enables spatially resolved analysis of unprepared samples in laser ablation experiments (cf. Fig. 3). The analytes released by laser bombardment are ionized directly with the SICRIT ion source, and the spatially resolved data are translated into two-dimensional images. This provides information, for example, on the distribution of active ingredients in tablets.Sabrina K. I. Funke, Valérie A. Brückel, Markus Weber, Elias Lützen, Jan-Christoph Wolf, Christoph Haisch, Uwe Karst: ''Plug-and-play laser ablation-mass spectrometry for molecular imaging by means of dielectric barrier discharge ionization''. In: ''Analytica Chimica Acta''. Volume 1177, September 8, 2021, ISSN 0003-2670, p. 338770, doi:10.1016/j.aca.2021.338770.


Cell analysis

In combination with a flow cytometer, the SICRIT ion source also enables the analysis of individual cells. The separated cell is introduced into the mass spectrometer through the ion source, and the lysate released upon cell rupture is analyzed. More precisely, the molecules contained in the lysate (mostly lipids) are ionized.