A collision/reaction cell is a device used in
inductively coupled plasma mass spectrometry Inductively coupled plasma mass spectrometry (ICP-MS) is a type of mass spectrometry that uses an inductively coupled plasma to ionize the sample. It atomizes the sample and creates atomic and small polyatomic ions, which are then detected. It i ...
to remove interfering
ion An ion () is an atom or molecule with a net electrical charge. The charge of an electron is considered to be negative by convention and this charge is equal and opposite to the charge of a proton, which is considered to be positive by convent ...
s through ion/neutral reactions.


Dynamic reaction cell

The collision reaction cell, also known by the trade name "dynamic reaction cell" (DRC), was introduced by
Perkin-Elmer PerkinElmer, Inc., previously styled Perkin-Elmer, is an American global corporation that was founded in 1937 and originally focused on precision optics. Over the years it went into and out of several different businesses via acquisitions and di ...
on their Elan DRC (followed by Elan DRC II and Elan DRC-e) instrument. The dynamic reaction cell is a chamber placed before the traditional
quadrupole A quadrupole or quadrapole is one of a sequence of configurations of things like electric charge or current, or gravitational mass that can exist in ideal form, but it is usually just part of a multipole expansion of a more complex structure re ...
chamber of an ICP-MS device, for eliminating isobaric interferences. The chamber has a quadrupole and can be filled with reaction (or collision) gases (
ammonia Ammonia is an inorganic chemical compound of nitrogen and hydrogen with the chemical formula, formula . A Binary compounds of hydrogen, stable binary hydride and the simplest pnictogen hydride, ammonia is a colourless gas with a distinctive pu ...
,
methane Methane ( , ) is a chemical compound with the chemical formula (one carbon atom bonded to four hydrogen atoms). It is a group-14 hydride, the simplest alkane, and the main constituent of natural gas. The abundance of methane on Earth makes ...
,
oxygen Oxygen is a chemical element; it has chemical symbol, symbol O and atomic number 8. It is a member of the chalcogen group (periodic table), group in the periodic table, a highly reactivity (chemistry), reactive nonmetal (chemistry), non ...
or
hydrogen Hydrogen is a chemical element; it has chemical symbol, symbol H and atomic number 1. It is the lightest and abundance of the chemical elements, most abundant chemical element in the universe, constituting about 75% of all baryon, normal matter ...
), with one gas type at a time or a mixture of two of them, which reacts with the introduced sample, eliminating some of the interference. The DRC is characterized by to main parameters that can be modified: RPq (the corresponding q parameter from the
Mathieu equation In mathematics, Mathieu functions, sometimes called angular Mathieu functions, are solutions of Mathieu's differential equation : \frac + (a - 2q\cos(2x))y = 0, where are real-valued parameters. Since we may add to to change the sign of , i ...
) and RPa (the corresponding a parameter from the
Mathieu equation In mathematics, Mathieu functions, sometimes called angular Mathieu functions, are solutions of Mathieu's differential equation : \frac + (a - 2q\cos(2x))y = 0, where are real-valued parameters. Since we may add to to change the sign of , i ...
). These parameters refer to the voltage applied to the quadrupole rods and the gas flow of the reaction gas. Ammonia gas is typically chosen to mitigate the majority of interferences. However, for specific
isotope Isotopes are distinct nuclear species (or ''nuclides'') of the same chemical element. They have the same atomic number (number of protons in their Atomic nucleus, nuclei) and position in the periodic table (and hence belong to the same chemica ...
s, other gasses may be required for better results, or mathematical correction if no gas offers a satisfactory advantage.


Collisional reaction interface (CRI) or mini-Collision/Reaction Cell

The proprietary collisional reaction interface (CRI)I. Kalinitchenko, Patent Application under the Patents Cooperation Treaty WO 2004/012223 A1 used in the
Bruker Bruker Corporation is an American manufacturer of scientific instruments for molecular and materials research, as well as for industrial and applied analysis. It is headquartered in Billerica, Massachusetts, and is the publicly traded parent comp ...
ICP-MS Aurora M90 destroys interfering ions. These ions are removed by injecting a collisional gas (He), or a reactive gas (H2), or a mixture of the two, directly into the plasma as it flows through the skimmer cone and/or the sampler cone. Supplying the reactive/collisional gas into the tip of the skimmer cone induces extra collisions and reactions that destroy polyatomic ions in the passing plasma. Fundamentally CRI is a mini- Collision/Reaction Cell installed in front of the parabolic Ion Mirror optics.


Axial field technology

Axial field technology (AFT) is a patented improvement of DRC made by Perkin-Elmer, which consists in two supplementary rods placed in the DRC cell, smaller than normal quadrupole's rods, with the purpose of "pushing" the ions faster to the exit by generating a supplementary
electric potential Electric potential (also called the ''electric field potential'', potential drop, the electrostatic potential) is defined as electric potential energy per unit of electric charge. More precisely, electric potential is the amount of work (physic ...
, minimizing the time needed for the gas to be in the DRC and improving analysis speed. The suplimetary potential of the AFT rods does not contribute significantly to the global energy, but drastically improve ion passage time.


Collision cell technology with kinetic energy discrimination

Thermo Scientific's XSeries2 instrument utilizes a collision/reaction cell for interference removal, consisting of a non-consumable hexapole and chicane ion deflector, which takes the ion beam off-axis and leads to low instrument backgrounds of <0.5 integrated counts per second (icps) at vacant masses such as 5 and 220. This hexapole is inherently part of the Thermo lens system and is present in the ion path, regardless of the use of the collision cell. The collision/reaction gas mixtures can be 1% NH3 in He, 7% H2 in He and 100% H2, where the NH3 and H2 are reactive gasses and the He is a collisional gas. The 3rd generation cell utilizes
kinetic energy In physics, the kinetic energy of an object is the form of energy that it possesses due to its motion. In classical mechanics, the kinetic energy of a non-rotating object of mass ''m'' traveling at a speed ''v'' is \fracmv^2.Resnick, Rober ...
discrimination, which employs running the quadrupole bias slightly less negative (more positive) than the hexapole bias. Polyatomic ions generated within the plasma can have larger atomic radii than
analyte An analyte, component (in clinical chemistry), titrand (in titrations), or chemical species is a substance or chemical constituent that is of interest in an analytical procedure. The remainder of the sample is called the matrix. The procedure ...
ions of similar mass, i.e. the interferent NaAr+ (mass 63) is larger than the analyte Cu+ (mass 63). Thus, when using a collisional/reactive gas mixture, these larger species undergo more collisions/reactions in the cell, in which they lose increasingly more energy, and are then excluded from the quadrupole mass filter by the kinetic energy barrier.


Octopole reaction system

Another implementation of this type of interference removal is an octopole (instead of a quadrupole) collision cell, implemented by
Agilent Agilent Technologies, Inc. is an American global company headquartered in Santa Clara, California, that provides instruments, software, services, and consumables for laboratories. Agilent was established in 1999 as a spin-off from Hewlett-Packar ...
's 7500 series. The octopole reaction system (ORS)) uses only helium or hydrogen and the volume of the cell is smaller than that of a DRC. The small molecules of helium and hydrogen collide with the large, unwanted polyatomic ions formed in the plasma and break them up into other ions that can be separated in the quadrupole mass analyser. However, unlike the DRC the OCR system is based only on collision reactions and not on chemical reactions.