Vastly undersampled Isotropic Projection Reconstruction
   HOME

TheInfoList



OR:

Phase contrast magnetic resonance imaging (PC-MRI) is a specific type of magnetic resonance imaging used primarily to determine flow
velocities Velocity is the directional speed of an object in motion as an indication of its rate of change in position as observed from a particular frame of reference and as measured by a particular standard of time (e.g. northbound). Velocity is a ...
. PC-MRI can be considered a method of Magnetic Resonance Velocimetry. It also provides a method of
magnetic resonance angiography Magnetic resonance angiography (MRA) is a group of techniques based on magnetic resonance imaging (MRI) to image blood vessels. Magnetic resonance angiography is used to generate images of arteries (and less commonly veins) in order to evaluate ...
. Since modern PC-MRI is typically time-resolved, it provides a means of 4D imaging (three spatial
dimension In physics and mathematics, the dimension of a mathematical space (or object) is informally defined as the minimum number of coordinates needed to specify any point within it. Thus, a line has a dimension of one (1D) because only one coor ...
s plus time).


How it Works

Atom Every atom is composed of a nucleus and one or more electrons bound to the nucleus. The nucleus is made of one or more protons and a number of neutrons. Only the most common variety of hydrogen has no neutrons. Every solid, liquid, gas, ...
s with an odd number of protons or neutrons have a randomly aligned angular spin momentum. When placed in a strong magnetic field, some of these spins align with the axis of the external field, which causes a net '
longitudinal Longitudinal is a geometric term of location which may refer to: * Longitude ** Line of longitude, also called a meridian * Longitudinal engine, an internal combustion engine in which the crankshaft is oriented along the long axis of the vehicl ...
' magnetization. These spins
precess Precession is a change in the orientation of the rotational axis of a rotating body. In an appropriate reference frame it can be defined as a change in the first Euler angle, whereas the third Euler angle defines the rotation itself. In othe ...
about the axis of the external field at a
frequency Frequency is the number of occurrences of a repeating event per unit of time. It is also occasionally referred to as ''temporal frequency'' for clarity, and is distinct from ''angular frequency''. Frequency is measured in hertz (Hz) which is eq ...
proportional to the strength of that field. Then, energy is added to the system through a
Radio frequency Radio frequency (RF) is the oscillation rate of an alternating electric current or voltage or of a magnetic, electric or electromagnetic field or mechanical system in the frequency range from around to around . This is roughly between the ...
(RF) pulse to 'excite' the spins, changing the axis that the spins precess about. These spins can then be observed by receiver coils (
Radiofrequency coil Radiofrequency coils (RF coils) are the receivers, and sometimes also the transmitters, of radiofrequency (RF) signals in equipment used in magnetic resonance imaging (MRI). The MR signal in MRI is produced by the process of resonance, which is the ...
s) using Faraday's law of induction. Different tissues respond to the added energy in different ways, and imaging parameters can be adjusted to highlight desired tissues. All of these spins have a phase that is dependent on the atom's velocity.
Phase shift In physics and mathematics, the phase of a periodic function F of some real variable t (such as time) is an angle-like quantity representing the fraction of the cycle covered up to t. It is denoted \phi(t) and expressed in such a scale that it ...
(\phi) of a spin is a function of the gradient field \mathbf(t): : \phi = \gamma \int_0^t B_0 + \mathbf(\tau) \cdot \mathbf(\tau) d\tau where \gamma is the Gyromagnetic ratio and \mathbf is defined as: : \mathbf (\tau) = \mathbf r_0 + \mathbf v_r \tau + \frac \mathbf a_r \tau^2 + \ldots , \mathbf r_0 is the initial position of the spin, \mathbf v_r is the spin velocity, and \mathbf a_r is the spin acceleration. If we only consider static spins and spins in the x-direction, we can rewrite equation for phase shift as: : \phi = \gamma x_0 \int_0^t G_x(\tau) d \tau + \gamma v_x \int_0^t G_x (\tau) \tau d \tau + \gamma \frac \int_0^t G_x ( \tau ) \tau^2 d \tau + \ldots We then assume that
acceleration In mechanics, acceleration is the rate of change of the velocity of an object with respect to time. Accelerations are vector quantities (in that they have magnitude and direction). The orientation of an object's acceleration is given by t ...
and higher order terms are negligible to simplify the expression for phase to: : \phi = \gamma (x_0 M_0 + v_x M_1) where M_0 is the zeroth moment of the x-gradient and M_1 is the first moment of the x gradient. If we take two different acquisitions with applied magnetic gradients that are the opposite of each other (bipolar gradients), we can add the results of the two acquisitions together to calculate a change in phase that is dependent on gradient: : \Delta \phi = v (\gamma \Delta M_1) where \Delta M_1 = 2 M_1 . The phase shift is measured and converted to a velocity according to the following equation: : v = \frac \Delta \phi where v_ is the maximum velocity that can be recorded and \Delta \phi is the recorded phase shift. The choice of v_ defines range of velocities visible, known as the ‘dynamic range’. A choice of v_ below the maximum velocity in the slice will induce aliasing in the image where a velocity just greater than v_ will be incorrectly calculated as moving in the opposite direction. However, there is a direct trade-off between the maximum velocity that can be encoded and the signal-to-noise ratio of the velocity measurements. This can be described by: : SNR_v = \frac \frac SNR where SNR is the signal-to-noise ratio of the image (which depends on the magnetic field of the scanner, the voxel volume, and the acquisition time of the scan). For an example, setting a ‘low’ v_ (below the maximum velocity expected in the scan) will allow for better visualization of slower velocities (better SNR), but any higher velocities will alias to an incorrect value. Setting a ‘high’ v_ (above the maximum velocity expected in the scan) will allow for the proper velocity quantification, but the larger dynamic range will obscure the smaller velocity features as well as decrease SNR. Therefore, the setting of v_ will be application dependent and care must be exercised in the selection. In order to further allow for proper velocity quantification, especially in clinical applications where the velocity dynamic range of flow is high (e.g. blood flow velocities in vessels across the thoracoabdominal cavity), a dual-echo PC-MRI (DEPC) method with dual velocity encoding in the same repetition time has been developed. The DEPC method does not only allow for proper velocity quantification, but also reduces the total acquisition time (especially when applied to 4D flow imaging) compared to a single-echo single-v_ PC-MRI acquisition carried out at two separate v_ values. To allow for more flexibility in selecting v_, instantaneous phase (phase unwrapping) can be used to increase both dynamic range and SNR.


Encoding Methods

When each dimension of velocity is calculated based on acquisitions from oppositely applied gradients, this is known as a six-point method. However, more efficient methods are also used. Two are described here:


Simple Four-point Method

Four sets of encoding gradients are used. The first is a reference and applies a negative moment in x, y, and z. The next applies a positive moment in x, and a negative moment in y and z. The third applies a positive moment in y, and a negative moment in x and z. And the last applies a positive moment in z, and a negative moment in x and y. Then, the velocities can be solved based on the phase information from the corresponding phase encodes as follows: : \hat_x = \frac : \hat_y = \frac : \hat_z = \frac


Balanced Four-Point Method

The balanced four-point method also includes four sets of encoding gradients. The first is the same as in the simple four-point method with negative gradients applied in all directions. The second has a negative moment in x, and a positive moment in y and z. The third has a negative moment in y, and a positive moment in x and z. The last has a negative moment in z and a positive moment in x and y. This gives us the following system of equations: : \phi_2 - \phi_1 = \phi_x + \phi_y : \phi_3 - \phi_1 = \phi_x + \phi_z : \phi_4 - \phi_1 = \phi_y + \phi_z Then, the velocities can be calculated: : \hat_x = \frac : \hat_y = \frac : \hat_z = \frac


Retrospective Cardiac and Respiratory Gating

For medical imaging, in order to get highly resolved scans in 3D space and time without motion artifacts from the heart or lungs, retrospective cardiac gating and respiratory compensation are employed. Beginning with cardiac gating, the patient's ECG signal is recorded throughout the imaging process. Similarly, the patient's respiratory patterns can be tracked throughout the scan. After the scan, the continuously collected data in K-space (magnetic resonance imaging), k-space (temporary image space) can be assigned accordingly to match-up with the timing of the heart beat and lung motion of the patient. This means that these scans are cardiac-averaged so the measured blood velocities are an average over multiple cardiac cycles.


Applications

Phase contrast MRI is one of the main techniques for
magnetic resonance angiography Magnetic resonance angiography (MRA) is a group of techniques based on magnetic resonance imaging (MRI) to image blood vessels. Magnetic resonance angiography is used to generate images of arteries (and less commonly veins) in order to evaluate ...
(MRA). This is used to generate images of arteries (and less commonly veins) in order to evaluate them for stenosis (abnormal narrowing), Vascular occlusion, occlusions, aneurysms (vessel wall dilatations, at risk of rupture) or other abnormalities. MRA is often used to evaluate the arteries of the neck and brain, the thoracic and abdominal aorta, the renal arteries, and the legs (the latter exam is often referred to as a "run-off").


Limitations

In particular, a few limitations of PC-MRI are of importance for the measured velocities: * Partial volume (imaging), Partial volume effects (when a voxel contains the boundary between static and moving materials) can overestimate phase leading to inaccurate velocities at the interface between materials or tissues. * Intravoxel phase dispersion (when velocities within a pixel are heterogeneous or in areas of turbulent flow) can produce a resultant phase that does not resolve the flow features accurately. * Assuming that acceleration and higher orders of motion are negligible can be inaccurate depending on the flow field. * Displacement artifacts (also known as misregistration and oblique flow artifacts) occur when there is a time difference between the phase and frequency encoding. These artifacts are highest when the flow direction is within the slice plane (most prominent in the heart and aorta for biological flows)


Vastly undersampled Isotropic Projection Reconstruction (VIPR)

A ''Vastly undersampled Isotropic Projection Reconstruction'' (VIPR) is a radially acquired MRI sequence which results in high-resolution MRA with significantly reduced scan times, and without the need for breath-holding.Page 602
in:


References

{{reflist Magnetic resonance imaging