Magnetic resonance imaging (MRI) is in general a safe technique, although injuries may occur as a result of failed safety procedures or human error. During the last 150 years, thousands of papers focusing on the effects or side effects of magnetic or radiofrequency fields have been published. They can be categorized as ''incidental'' and ''physiological''.
Contraindications In medicine, a contraindication is a condition that serves as a reason not to take a certain medical treatment due to the harm that it would cause the patient. Contraindication is the opposite of indication, which is a reason to use a certain tre ...
to MRI include most
cochlear implant A cochlear implant (CI) is a surgically implanted neuroprosthesis that provides a person who has moderate-to-profound sensorineural hearing loss with sound perception. With the help of therapy, cochlear implants may allow for improved speech und ...
s and
cardiac pacemakers 350px, Image showing the cardiac pacemaker or SA node, the primary pacemaker within the electrical_conduction_system_of_the_heart">SA_node,_the_primary_pacemaker_within_the_electrical_conduction_system_of_the_heart. The_muscle_contraction.htm ...
,
shrapnel Shrapnel may refer to: Military * Shrapnel shell, explosive artillery munitions, generally for anti-personnel use * Shrapnel (fragment), a hard loose material Popular culture * ''Shrapnel'' (Radical Comics) * ''Shrapnel'', a game by Adam C ...
and metallic
foreign bodies A foreign body (FB) is any object originating outside the body of an organism. In machinery, it can mean any unwanted intruding object. Most references to foreign bodies involve propulsion through natural orifices into hollow organs. Foreign bo ...
in the
eyes Eyes are organs of the visual system. They provide living organisms with vision, the ability to receive and process visual detail, as well as enabling several photo response functions that are independent of vision. Eyes detect light and conve ...
. The safety of MRI during the first trimester of pregnancy is uncertain, but it may be preferable to other options. Since MRI does not use any
ionizing radiation Ionizing radiation (or ionising radiation), including nuclear radiation, consists of subatomic particles or electromagnetic waves that have sufficient energy to ionize atoms or molecules by detaching electrons from them. Some particles can travel ...
, its use generally is favored in preference to CT when either modality could yield the same information. (In certain cases, MRI is not preferred as it may be more expensive, time-consuming and
claustrophobia Claustrophobia is the fear of confined spaces. It can be triggered by many situations or stimuli, including elevators, especially when crowded to capacity, windowless rooms, and hotel rooms with closed doors and sealed windows. Even bedrooms wit ...
-exacerbating).


Structure and certification

In an effort to standardize the roles and responsibilities of MRI professionals, an international consensus document, written and endorsed by major MRI and medical physics professional societies from around the globe, has been published formally. The document outlines specific responsibilities for the following positions: * ''MR Medical Director / Research Director (MRMD)'' – This individual is the supervising physician who has oversight responsibility for the safe use of MRI services. * ''MR Safety Officer (MRSO)'' – Roughly analogous to a radiation safety officer, the MRSO acts on behalf of, and on the instruction of, the MRMD to execute safety procedures and practices at the point of care. * ''MR Safety Expert (MRSE)'' – This individual serves in a consulting role to both the MRMD and MRSO, assisting in the investigation of safety questions that may include the need for extrapolation, interpolation, or quantification to approximate the risk of a specific study. Th
American Board of Magnetic Resonance Safety
(ABMRS) provides testing and board certification for each of the three positions, MRMD, MRSO, and MRSE. As most MRI accidents and injuries are directly attributable to decisions at the point of care, testing and certification of MRI professionals seeks to reduce the rates of MRI accidents and improve patient safety through the establishment of safety competency levels for MRI professionals.


Implants

All patients are reviewed for contraindications prior to MRI scanning. Medical devices and implants are categorized as MR Safe, MR Conditional or MR Unsafe: * ''MR-Safe'' – The device or implant is completely non-magnetic, non-electrically conductive, and non-RF reactive, eliminating all of the primary potential threats during an MRI procedure. * ''MR-Conditional'' – A device or implant that may contain magnetic, electrically conductive, or RF-reactive components that is safe for operations in proximity to the MRI, provided the conditions for safe operation are defined and observed (such as 'tested safe to 1.5 teslas' or 'safe in magnetic fields below 500
gauss Johann Carl Friedrich Gauss (; german: Gauß ; la, Carolus Fridericus Gauss; 30 April 177723 February 1855) was a German mathematician and physicist who made significant contributions to many fields in mathematics and science. Sometimes refer ...
in strength'). * ''MR-Unsafe'' – Objects that are significantly ferromagnetic and pose a clear and direct threat to persons and equipment within the magnet room. The MRI environment may cause harm in patients with MR-Unsafe devices such as
cochlear implant A cochlear implant (CI) is a surgically implanted neuroprosthesis that provides a person who has moderate-to-profound sensorineural hearing loss with sound perception. With the help of therapy, cochlear implants may allow for improved speech und ...
s, aneurysm clips, and many permanent pacemakers. In November 1992, a patient with an undisclosed cerebral aneurysm clip was reported to have died shortly after an MRI exam. Several deaths have been reported in patients with pacemakers who have undergone MRI scanning without appropriate precautions. Increasingly, MR-conditional pacemakers are available for selected patients.
Ferromagnetic Ferromagnetism is a property of certain materials (such as iron) which results in a large observed magnetic permeability, and in many cases a large magnetic coercivity allowing the material to form a permanent magnet. Ferromagnetic materials ...
foreign bodies such as
shell Shell may refer to: Architecture and design * Shell (structure), a thin structure ** Concrete shell, a thin shell of concrete, usually with no interior columns or exterior buttresses ** Thin-shell structure Science Biology * Seashell, a hard o ...
fragments, or metallic implants such as surgical prostheses and ferromagnetic
aneurysm An aneurysm is an outward bulging, likened to a bubble or balloon, caused by a localized, abnormal, weak spot on a blood vessel wall. Aneurysms may be a result of a hereditary condition or an acquired disease. Aneurysms can also be a nidus (s ...
clips also are potential risks. Interaction of the magnetic and radio frequency fields with such objects may lead to heating or torque of the object during an MRI. MRI is contraindicated in those suspected with metallic foreign body in the eye. MRI maybe considered if there is strong suspicion of non-metallic foreign body.
Titanium Titanium is a chemical element with the symbol Ti and atomic number 22. Found in nature only as an oxide, it can be reduced to produce a lustrous transition metal with a silver color, low density, and high strength, resistant to corrosion in ...
and its alloys are safe from attraction and torque forces produced by the magnetic field, although there may be some risks associated with Lenz effect forces acting on titanium implants in sensitive areas within the subject, such as
stapes The ''stapes'' or stirrup is a bone in the middle ear of humans and other animals which is involved in the conduction of sound vibrations to the inner ear. This bone is connected to the oval window by its annular ligament, which allows the foo ...
implants in the inner ear. Intrauterine devices with copper are generally safe in MRI, but may become dislodged or even expelled, and it is therefore recommended to check the location of the IUD both before and after MRI. Other implants that are contraindicated in MRI includes: magnetic dental implants, tissue expander, artificial limb, hearing aid, catheters with metallic components such as Swan-Ganz catheter and piercing. However, tooth amalgam is not contraindicated in MRI.


Risk of implant heating under MRI

Titanium and its alloys can heat from the radiofrequency field, as well as the switched gradient field (due to Faraday's law of magnetic induction). The amount of heating that takes place has a number of contributing factors: Injuries have been reported by this heating of metallic implants:


Projectile risk

The very high strength of the magnetic field may cause
projectile A projectile is an object that is propelled by the application of an external force and then moves freely under the influence of gravity and air resistance. Although any objects in motion through space are projectiles, they are commonly found in ...
effect (or "missile-effect") accidents, where ferromagnetic objects are attracted to the center of the magnet.
Pennsylvania Pennsylvania (; ( Pennsylvania Dutch: )), officially the Commonwealth of Pennsylvania, is a state spanning the Mid-Atlantic, Northeastern, Appalachian, and Great Lakes regions of the United States. It borders Delaware to its southeast, ...
reported 27 cases of objects becoming projectiles in the MRI environment between 2004 and 2008. There have been incidents of injury and death.Randal C. Archibold,
Hospital Details Failures Leading to M.R.I. Fatality
''
The New York Times ''The New York Times'' (''the Times'', ''NYT'', or the Gray Lady) is a daily newspaper based in New York City with a worldwide readership reported in 2020 to comprise a declining 840,000 paid print subscribers, and a growing 6 million paid ...
'', August 22, 2001
Donald G. McNeil Jr,
M.R.I.'s Strong Magnets Cited in Accidents
" ''
The New York Times ''The New York Times'' (''the Times'', ''NYT'', or the Gray Lady) is a daily newspaper based in New York City with a worldwide readership reported in 2020 to comprise a declining 840,000 paid print subscribers, and a growing 6 million paid ...
'', August 19, 2005.
In one case, a six-year-old boy died in July 2001, during an MRI exam at the
Westchester Medical Center Westchester Medical Center University Hospital (WMC), formerly Grasslands Hospital, is an 895-bed Regional Trauma Center providing health services to residents of the Hudson Valley, northern New Jersey, and southern Connecticut. It is known for h ...
, New York, after a metal oxygen tank was pulled across the room and crushed the child's head. To reduce the risk of projectile accidents, ferromagnetic objects and devices are typically prohibited near the MRI scanner, and patients undergoing MRI examinations must remove all metallic objects, often by changing into a gown or scrubs. Some radiology departments use ferromagnetic detection devices to ensure that no ferromagnetic objects enter the scanner room.


MRI-EEG

In research settings, structural MRI or functional MRI (fMRI) may be combined with EEG (
electroencephalography Electroencephalography (EEG) is a method to record an electrogram of the spontaneous electrical activity of the brain. The biosignals detected by EEG have been shown to represent the postsynaptic potentials of pyramidal neurons in the neocortex ...
) under the condition that the EEG equipment is MR-compatible. Although EEG equipment (electrodes, amplifiers, and peripherals) are either approved for research or clinical use, the same MR Safe, MR Conditional and MR Unsafe terminology applies. With the growth of the use of MR technology, the U.S. Food & Drug Administration DArecognized the need for a consensus on standards of practice, and the FDA sought out
ASTM International ASTM International, formerly known as American Society for Testing and Materials, is an international standards organization that develops and publishes voluntary consensus technical standards for a wide range of materials, products, systems, ...
STMto achieve them. Committee F04 of ASTM developed F2503, Standard Practice for Marking Medical Devices and Other Items for Safety in the Magnetic Resonance Environment.


Genotoxic effects

There is no proven risk of biological harm from any aspect of an MRI scan, including very powerful static magnetic fields, gradient magnetic fields, or radio frequency waves. Some studies have suggested possible
genotoxic Genotoxicity is the property of chemical agents that damage the genetic information within a cell causing mutations, which may lead to cancer. While genotoxicity is often confused with mutagenicity, all mutagens are genotoxic, but some genotoxic s ...
(i.e., potentially carcinogenic) effects of MRI scanning through micronuclei induction and DNA double strand breaks in vivo and in vitro, however, in most, if not all cases, others have been unable to repeat or validate the results of these studies, and the majority of research shows no genotoxic, or otherwise harmful, effects caused by any part of MRI. A recent study confirmed that MRI using some of the most potentially-risky parameters tested to date (7-tesla static magnetic field, 70 mT/m gradient magnetic field, and maximum strength radio frequency waves) did not cause any DNA damage ''in vitro''.


Peripheral nerve stimulation

The rapid switching on and off of the magnetic field gradients is capable of causing nerve stimulation. Volunteers report a twitching sensation when exposed to rapidly switched fields, particularly in their extremities. The reason the peripheral nerves are stimulated is that the changing field increases with distance from the center of the gradient coils (which more or less coincides with the center of the magnet). Although PNS was not a problem for the slow, weak gradients used in the early days of MRI, the strong, rapidly switched gradients used in techniques such as EPI, fMRI, diffusion MRI, etc. are capable of inducing PNS. American and European regulatory agencies insist that manufacturers stay below specified ''d''B/''dt'' limits (''d''B/''dt'' is the change in magnetic field strength per unit time), or else prove that no PNS is induced for any imaging sequence. As a result of ''d''B/''dt'' limitation, commercial MRI systems cannot use the full rated power of their gradient amplifiers.


Heating caused by absorption of radio waves

Every MRI scanner has a powerful radio transmitter that generates the electromagnetic field that excites the spins. If the body absorbs the energy, heating occurs. For this reason, the transmitter rate at which energy is absorbed by the body must be limited (see
Specific absorption rate Specific absorption rate (SAR) is a measure of the rate at which energy is absorbed per unit mass by a human body when exposed to a radio frequency (RF) electromagnetic field. It can also refer to absorption of other forms of energy by tissue, inc ...
). It has been claimed that tattoos made with iron-containing dyes may lead to burns on the subject's body. Cosmetics are very unlikely to undergo heating, as well as body lotions, since the outcome of the reactions between those with the radio waves is unknown. The best option for clothing is 100% cotton. There are several positions strictly forbidden during measurement such as crossing arms and legs, and the patient's body may not create loops of any kind for the RF during the measurement.


Acoustic noise

Switching of field gradients causes a change in the Lorentz force experienced by the gradient coils, producing minute expansions and contractions of the coil. As the switching typically is in the audible frequency range, the resulting vibration produces loud noises (clicking, banging or beeping). This behaviour, of sound being generated by the vibration of the conducting components, is described as a coupled acousto-magneto-mechanical system, solutions to which provide useful insight to the behaviour of the scanners. This is most marked with high-field machines,"The Evolution of Magnetic Resonance Imaging: 3T MRI in Clinical Applications"
, Terry Duggan-Jahns, www.eradimaging.com
and rapid-imaging techniques in which sound pressure levels may reach 120
dB(A) A-weighting is the most commonly used of a family of curves defined in the International standard IEC 61672:2003 and various national standards relating to the measurement of sound pressure level. A-weighting is applied to instrument-measured s ...
(equivalent to a jet engine at take-off), and therefore, appropriate ear protection is essential for anyone inside the MRI scanner room during the examination.
The Open University The Open University (OU) is a British public research university and the largest university in the United Kingdom by number of students. The majority of the OU's undergraduate students are based in the United Kingdom and principally study o ...
2007: ''Understanding Cardiovascular Diseases'', course book for the lesson
SK121 Understanding cardiovascular diseases
', printed by
Cambridge University Press Cambridge University Press is the university press of the University of Cambridge. Granted letters patent by Henry VIII of England, King Henry VIII in 1534, it is the oldest university press A university press is an academic publishing hou ...
, (can be found a
OUW
, pages 220 and 224.
Radio frequency in itself does not cause audible noises (at least for human beings), since modern systems are using frequencies of 8.5 MHz (0.2 T system) or higher.


Cryogens

As described in the
Physics of magnetic resonance imaging The physics of magnetic resonance imaging (MRI) concerns fundamental physical considerations of MRI techniques and technological aspects of MRI devices. MRI is a medical imaging technique mostly used in radiology and nuclear medicine in order t ...
article, many MRI scanners rely on cryogenic liquids to enable the superconducting capabilities of the electromagnetic coils within. Although the cryogenic liquids used are non-toxic, their physical properties present specific hazards. An unintentional shut-down of a
superconducting electromagnet A superconducting magnet is an electromagnet made from coils of superconducting wire. They must be cooled to cryogenic temperatures during operation. In its superconducting state the wire has no electrical resistance and therefore can conduct mu ...
, an event known as "quench", involves the rapid boiling of liquid
helium Helium (from el, ἥλιος, helios, lit=sun) is a chemical element with the symbol He and atomic number 2. It is a colorless, odorless, tasteless, non-toxic, inert, monatomic gas and the first in the noble gas group in the periodic table. ...
from the device. If the rapidly expanding helium cannot be dissipated through an external vent, sometimes referred to as a 'quench pipe', it may be released into the scanner room where it may cause displacement of the oxygen and present a risk of
asphyxiation Asphyxia or asphyxiation is a condition of deficient supply of oxygen to the body which arises from abnormal breathing. Asphyxia causes generalized hypoxia, which affects primarily the tissues and organs. There are many circumstances that can i ...
. Oxygen deficiency monitors usually are used as a safety precaution. Liquid helium, the most commonly used cryogen in MRI, undergoes near explosive expansion as it changes from a liquid to gaseous state. The use of an oxygen monitor is important to ensure that oxygen levels are safe for patients and physicians. Rooms built for superconducting MRI equipment should be equipped with pressure relief mechanisms and an exhaust fan, in addition to the required quench pipe. Because a quench results in rapid loss of
cryogen In physics, cryogenics is the production and behaviour of materials at very low temperatures. The 13th IIR International Congress of Refrigeration (held in Washington DC in 1971) endorsed a universal definition of “cryogenics” and “cr ...
s from the magnet, recommissioning the magnet is expensive and time-consuming. Spontaneous quenches are uncommon, but a quench also may be triggered by an equipment malfunction, an improper cryogen fill technique, contaminants inside the cryostat, or extreme magnetic or vibrational disturbances.


Pregnancy

No effects of MRI on the fetus have been demonstrated. As opposed to many other forms of
medical imaging in pregnancy Medical imaging in pregnancy may be indicated because of pregnancy complications, intercurrent diseases or routine prenatal care. Options Options for medical imaging in pregnancy include the following: *Magnetic resonance imaging (MRI) without ...
, MRI avoids the use of
ionizing radiation Ionizing radiation (or ionising radiation), including nuclear radiation, consists of subatomic particles or electromagnetic waves that have sufficient energy to ionize atoms or molecules by detaching electrons from them. Some particles can travel ...
, to which the fetus is particularly sensitive. As a precaution, however, many guidelines recommend pregnant women only undergo MRI when essential, especially during the first trimester. The concerns in pregnancy are the same as for MRI in general, but the fetus may be more sensitive to the effects—particularly to heating and to noise. The use of
gadolinium Gadolinium is a chemical element with the symbol Gd and atomic number 64. Gadolinium is a silvery-white metal when oxidation is removed. It is only slightly malleable and is a ductile rare-earth element. Gadolinium reacts with atmospheric oxygen ...
-based contrast media in pregnancy is an off-label indication and may be administered only in the lowest dose required to provide essential diagnostic information. Despite these concerns, MRI is rapidly growing in importance as a way of diagnosing and monitoring
congenital defect A birth defect, also known as a congenital disorder, is an abnormal condition that is present at birth regardless of its cause. Birth defects may result in disabilities that may be physical, intellectual, or developmental. The disabilities can r ...
s of the fetus because it is able to provide more diagnostic information than
ultrasound Ultrasound is sound waves with frequency, frequencies higher than the upper audible limit of human hearing range, hearing. Ultrasound is not different from "normal" (audible) sound in its physical properties, except that humans cannot hea ...
and it lacks the ionizing radiation of CT. MRI without contrast agents is the imaging mode of choice for pre-surgical, in-utero diagnosis and evaluation of fetal tumors, primarily
teratoma A teratoma is a tumor made up of several different types of tissue, such as hair, muscle, teeth, or bone. Teratomata typically form in the ovary, testicle, or coccyx. Symptoms Symptoms may be minimal if the tumor is small. A testicular ter ...
s, facilitating open
fetal surgery Fetal surgery also known as antenatal surgery, prenatal surgery, is a growing branch of maternal-fetal medicine that covers any of a broad range of surgical techniques that are used to treat congenital abnormalities in fetuses who are still i ...
, other
fetal intervention Fetal surgery also known as antenatal surgery, prenatal surgery, is a growing branch of maternal-fetal medicine that covers any of a broad range of surgical techniques that are used to treat congenital abnormalities in fetuses who are still in ...
s, and planning for procedures (such as the
EXIT procedure The EXIT procedure, or ex utero intrapartum treatment procedure, is a specialized surgical delivery procedure used to deliver babies who have airway compression. Causes of airway compression in newborn babies result from a number of rare congen ...
) to safely deliver and treat babies whose defects would otherwise be fatal.


Claustrophobia and discomfort

Although painless, MRI scans may be unpleasant for those who are claustrophobic or otherwise uncomfortable with the imaging device surrounding them. Older closed bore MRI systems have a fairly long tube or tunnel. The part of the body being imaged must lie at the center of the magnet, which is at the absolute center of the tunnel. Because scan times on these older scanners may be long (occasionally up to 40 minutes for the entire procedure), people with even mild claustrophobia are sometimes unable to tolerate an MRI scan without management. Some modern scanners have larger bores (up to 70 cm) and scan times are shorter. A 1.5 T wide short bore scanner increases the examination success rate in patients with claustrophobia and substantially reduces the need for anesthesia-assisted MRI examinations even when claustrophobia is severe. Alternative scanner designs, such as open or upright systems, may be helpful where these are available. Although open scanners have increased in popularity, they produce inferior scan quality because they operate at lower magnetic fields than closed scanners. Commercial 1.5- tesla open systems have become available recently, however, providing much better image quality than previous lower field strength open models. Mirror glasses may be used to help create the illusion of openness. The mirrors are angled at 45 degrees, allowing the patient to look down their body and out the end of the imaging area. The appearance is of an open tube pointing upward (as seen when lying in the imaging area). Even though one is able to see around the glasses and the proximity of the device is very evident, this illusion is quite persuasive and relieves the claustrophobic feeling. For young children who cannot hold still or would be frightened during the examination, chemical sedation or general anesthesia are the norm. Some hospitals encourage children to pretend the MRI machine is a spaceship or other adventure. Certain hospitals with Children's wards have decorated scanners for this purpose, such as that at the
Boston Children's Hospital Boston Children's Hospital formerly known as Children's Hospital Boston until 2012 is a nationally ranked, freestanding acute care children's hospital located in Boston, Massachusetts, adjacent both to its teaching affiliate, Harvard Medical Scho ...
, which operates a scanner with a special casing designed to resemble a
sandcastle Sand art is the practice of modelling sand into an artistic form, such as sand brushing, sand sculpting, sand painting, or creating sand bottles. A sandcastle is a type of sand sculpture resembling a miniature building, often a castle. The dr ...
. Obese patients and pregnant women may find the MRI machine a tight fit. Pregnant women in the third trimester also may have difficulty lying on their backs for an hour or more without moving.


MRI versus CT

MRI and computed tomography (CT) are complementary imaging technologies and each has advantages and limitations for particular applications. CT is more widely used than MRI in
OECD The Organisation for Economic Co-operation and Development (OECD; french: Organisation de coopération et de développement économiques, ''OCDE'') is an intergovernmental organisation with 38 member countries, founded in 1961 to stimulate e ...
countries with a mean of 132 vs. 46 exams per 1000 population performed respectively. A concern is the potential for CT to contribute to
radiation-induced cancer Exposure to ionizing radiation is known to increase the future incidence of cancer, particularly leukemia. The mechanism by which this occurs is well understood, but quantitative models predicting the level of risk remain controversial. The most wi ...
and in 2007 it was estimated that 0.4% of current cancers in the United States were due to CTs performed in the past, and that in the future this figure may rise to 1.5–2% based on historical rates of CT usage. An Australian study found that one in every 1800 CT scans was associated with an excess cancer. An advantage of MRI is that no ionizing radiation is used and so it is recommended over CT when either approach could yield the same diagnostic information. Although the cost of MRI has fallen, making it more competitive with CT, there are not many common imaging scenarios in which MRI can simply replace CT, however, this substitution has been suggested for the imaging of liver disease. The effect of low doses of radiation on carcinogenesis also are disputed. Although MRI is associated with biological effects, these have not been proven to cause measurable harm. Iodinated contrast medium is routinely used in CT and the main adverse events are anaphylactoid reactions and
nephrotoxicity Nephrotoxicity is toxicity in the kidneys. It is a poisonous effect of some substances, both toxic chemicals and medications, on kidney function. There are various forms, and some drugs may affect kidney function in more than one way. Nephrotoxin ...
. Commonly used
MRI contrast agent MRI contrast agents are contrast agents used to improve the visibility of internal body structures in magnetic resonance imaging (MRI). The most commonly used compounds for contrast enhancement are gadolinium-based. Such MRI contrast agents shorten ...
s have a good safety profile, but linear non-ionic agents in particular have been implicated in
nephrogenic systemic fibrosis Nephrogenic systemic fibrosis is a rare syndrome that involves fibrosis of skin, joints, eyes, and internal organs. NSF is caused by exposure to gadolinium in gadolinium-based MRI contrast agents (GBCAs) in patients with impaired kidney function ...
in patients with severely impaired renal function. MRI is contraindicated in the presence of MR-unsafe implants, and although these patients may be imaged with CT, beam hardening artefact from metallic devices, such as
pacemakers An artificial cardiac pacemaker (or artificial pacemaker, so as not to be confused with the natural cardiac pacemaker) or pacemaker is a medical device that generates electrical impulses delivered by electrodes to the chambers of the heart eit ...
and
implantable cardioverter-defibrillator An implantable cardioverter-defibrillator (ICD) or automated implantable cardioverter defibrillator (AICD) is a device implantable inside the body, able to perform defibrillation, and depending on the type, cardioversion and pacing of the h ...
s, also may affect image quality. MRI is a longer investigation than CT and an exam may take between 20 and 40 minutes depending on complexity.


Guidance

Safety issues, including the potential for biostimulation device interference, movement of ferromagnetic bodies, and incidental localized heating, have been addressed in the
American College of Radiology The American College of Radiology (ACR), founded in 1923, is a professional medical society representing nearly 40,000 diagnostic radiologists, radiation oncologists, interventional radiologists, nuclear medicine physicians and medical physicists. ...
's ''White Paper on MR Safety'', which originally was published in 2002 and expanded in 2004. The ''ACR White Paper on MR Safety'' has been rewritten and was released early in 2007 under the new title ''ACR Guidance Document for Safe MR Practices''. In December 2007, the Medicines and Healthcare products Regulatory Agency (MHRA), a UK healthcare regulatory body, issued their ''Safety Guidelines for Magnetic Resonance Imaging Equipment in Clinical Use''. In February 2008, the
Joint Commission The Joint Commission is a United States-based nonprofit tax-exempt 501(c) organization that accredits more than 22,000 US health care organizations and programs. The international branch accredits medical services from around the world. A majori ...
, a U.S. healthcare accrediting organization, issued a Sentinel Event Alert #38, their highest patient safety advisory, on MRI safety issues. In July 2008, the United States Veterans Administration, a federal governmental agency serving the healthcare needs of former military personnel, issued a substantial revision to their ''MRI Design Guide'', that includes physical and facility safety considerations.


The European Directive on electromagnetic fields

This Directive (2013/35/EU – electromagnetic fields) covers all known direct biophysical effects and indirect effects caused by electromagnetic fields within the EU and repealed the 2004/40/EC directive. The deadline for implementation of the new directive was 1 July 2016. Article 10 of the directive sets out the scope of the derogation for MRI, stating that the exposure limits may be exceeded during "the installation, testing, use, development, maintenance of or research related to magnetic resonance imaging (MRI) equipment for patients in the health sector, provided that certain conditions are met." Uncertainties remain regarding the scope and conditions of this derogation.


References

{{reflist Magnetic resonance imaging
Magnetic resonance imaging Magnetic resonance imaging (MRI) is a medical imaging technique used in radiology to form pictures of the anatomy and the physiological processes of the body. MRI scanners use strong magnetic fields, magnetic field gradients, and radio wave ...