HOME

TheInfoList



OR:

A helium mass spectrometer is an instrument commonly used to detect and locate small leaks. It was initially developed in the
Manhattan Project The Manhattan Project was a research and development undertaking during World War II that produced the first nuclear weapons. It was led by the United States with the support of the United Kingdom and Canada. From 1942 to 1946, the project w ...
during World War II to find extremely small leaks in the gas diffusion process of uranium enrichment plants. It typically uses a
vacuum chamber A vacuum chamber is a rigid enclosure from which air and other gases are removed by a vacuum pump. This results in a low-pressure environment within the chamber, commonly referred to as a vacuum. A vacuum environment allows researchers to con ...
in which a sealed container filled with helium is placed. Helium leaks out of the container, and the rate of the leak is detected by a
mass spectrometer Mass spectrometry (MS) is an analytical technique that is used to measure the mass-to-charge ratio of ions. The results are presented as a '' mass spectrum'', a plot of intensity as a function of the mass-to-charge ratio. Mass spectrometry is us ...
.


Detection technique

Helium is used as a
tracer Tracer may refer to: Science * Flow tracer, any fluid property used to track fluid motion * Fluorescent tracer, a substance such as 2-NBDG containing a fluorophore that is used for tracking purposes * Histochemical tracer, a substance used for ...
because it penetrates small leaks rapidly. Helium also has the properties of being non-toxic,
chemically inert In chemistry, the term chemically inert is used to describe a substance that is not chemically reactive. From a thermodynamic perspective, a substance is inert, or nonlabile, if it is thermodynamically unstable (positive standard Gibbs free en ...
and present in the atmosphere only in minute quantities (5  ppm). Typically a helium leak detector will be used to measure leaks in the range of 10 to 10  Pa· m· s. A flow of 10 Pa·m·s is about 0.006 ml per minute at
standard conditions for temperature and pressure Standard temperature and pressure (STP) are standard sets of conditions for experimental measurements to be established to allow comparisons to be made between different sets of data. The most used standards are those of the International Union ...
(STP). A flow of 10 Pa·m·s is about 0.003 ml per century at STP.


Types of leaks

Typically there are two types of leaks in the detection of helium as a ''
tracer Tracer may refer to: Science * Flow tracer, any fluid property used to track fluid motion * Fluorescent tracer, a substance such as 2-NBDG containing a fluorophore that is used for tracking purposes * Histochemical tracer, a substance used for ...
'' for leak detection: residual leak and virtual leak. A residual leak is a real leak due to an imperfect seal, a puncture, or some other hole in the system. A virtual leak is the semblance of a leak in a vacuum system caused by
outgassing Outgassing (sometimes called offgassing, particularly when in reference to indoor air quality) is the release of a gas that was dissolved, trapped, frozen, or absorbed in some material. Outgassing can include sublimation and evaporation (which ...
of chemicals trapped or adhered to the interior of a system that is actually sealed. As the gases are released into the chamber, they can create a
false positive A false positive is an error in binary classification in which a test result incorrectly indicates the presence of a condition (such as a disease when the disease is not present), while a false negative is the opposite error, where the test resul ...
indication of a residual leak in the system.


Uses

Helium mass spectrometer leak detectors are used in production line industries such as
refrigeration The term refrigeration refers to the process of removing heat from an enclosed space or substance for the purpose of lowering the temperature.International Dictionary of Refrigeration, http://dictionary.iifiir.org/search.phpASHRAE Terminology, ht ...
and
air conditioning Air conditioning, often abbreviated as A/C or AC, is the process of removing heat from an enclosed space to achieve a more comfortable interior environment (sometimes referred to as 'comfort cooling') and in some cases also strictly controlling ...
, automotive parts, carbonated beverage containers food packages and
aerosol An aerosol is a suspension of fine solid particles or liquid droplets in air or another gas. Aerosols can be natural or anthropogenic. Examples of natural aerosols are fog or mist, dust, forest exudates, and geyser steam. Examples of anthropogen ...
packaging, as well as in the manufacture of steam products,
gas bottle Bottled gas is a term used for substances which are gaseous at standard temperature and pressure (STP) and have been compressed and stored in carbon steel, stainless steel, aluminum, or composite bottles known as gas cylinders. Gas state in cy ...
s,
fire extinguisher A fire extinguisher is a handheld active fire protection device usually filled with a dry or wet chemical used to extinguish or control small fires, often in emergencies. It is not intended for use on an out-of-control fire, such as one which ha ...
s,
tire A tire (American English) or tyre (British English) is a ring-shaped component that surrounds a Rim (wheel), wheel's rim to transfer a vehicle's load from the axle through the wheel to the ground and to provide Traction (engineering), t ...
valves, and numerous other products including all
vacuum A vacuum is a space devoid of matter. The word is derived from the Latin adjective ''vacuus'' for "vacant" or " void". An approximation to such vacuum is a region with a gaseous pressure much less than atmospheric pressure. Physicists often ...
systems.


Test methods


Global helium spray

This method requires the part to be tested to be connected to a helium leak detector. The outer surface of the part to be tested will be located in some kind of a tent in which the helium concentration will be raised to 100% helium. If the part is small the vacuum system included in the leak testing instrument will be able to reach low enough pressure to allow for mass spectrometer operation. If the size of the part is too large, an additional vacuum pumping system may be required to reach low enough pressure in a reasonable length of time. Once operating pressure has been reached, the mass spectrometer can start its measuring operation. If leakage is encountered the small and "agile" molecules of helium will migrate through the cracks into the part. The vacuum system will carry any tracer gas molecule into the analyzer cell of the magnetic sector mass spectrometer. A signal will inform the operator of the value of the leakage encountered.


Local helium spray

This method is a small variation from the one above. It still requires the part to be tested to be connected to a helium leak detector. The outer surface of the part to be tested is sprayed with a localized stream of helium tracer gas. If the part is small the vacuum system included in the instrument will be able to reach low enough pressure to allow for mass spectrometer operation. If the size of the part is too large, an additional pumping system may be required to reach low enough pressure in a reasonable length of time. Once operating pressure has been reached, the mass spectrometer can start its measuring operation. If leakage is encountered the small and "agile" molecules of helium will migrate through the cracks into the part. The vacuum system will carry any tracer gas molecule into the analyzer cell of the magnetic sector mass spectrometer. A signal will inform the operator of the value of the leakage encountered. Thus correlation between maximum leakage signal and location of helium spray head will allow the operator to pinpoint the leaky area.


Helium charged vacuum test

In this case the part is pressurized (sometime this test is combined with a burst test, i.e. at 40 bar) with helium while sitting in a vacuum chamber. The vacuum chamber is connected to a vacuum pumping system and a leak detector. Once the vacuum has reached the mass spectrometer operating pressure, any helium leakage will be measured. This test method applies to a lot of components that will operate under pressure: airbag canisters, evaporators, condensers, high-voltage SF6 filled switchgear.


Partial vacuum method (ultra sniffer test)

In contrast to the Helium charged sniffer test, the partial vacuum method, the ultra sniffer test gas method (UST-method) uses the partial vacuum effect, so that the gas tightness of test sample can be detected at normal pressure with the same sensitivity as the helium charged vacuum test with helium gas helium. The method has a sensitivity of 10 Pa·m·s. Similar to the classical Helium charged sniffer test the test sample is enclosed in a bag, but in contrast to the classic method, the bag is exposed with a helium-free gas, so that the helium concentration inside the bag can reduced from 5·10 to 10 Pa·m·s. This sensitivity corresponds to a theoretical gas loss of 1 cm in 3000 years. The UST method can be used very economically for the ad hoc testing of test samples. The test system can be set up easily, with normal pneumatic items, such as valves and plastic hoses. For the embedding of the test samples, a simple plastic bag is sufficient. The UST method was also used for the leak testing of component of the fusion experiment Wendelstein 7-X in Germany.


Bombing test

This method applies to objects that are supposedly sealed. First the device under test will be exposed for an extended length of time to a high helium pressure in a "bombing" chamber. If the part is leaky, helium will be able to penetrate the device. Later the device will be placed in a vacuum chamber, connected to a vacuum pump and a mass spectrometer. The tiny amount of gas that entered the device under pressure will be released in the vacuum chamber and sent to the mass spectrometer where the leak rate will be measured. This test method applies to implantable medical devices, crystal oscillator, saw filter devices. This method is not able to detect a massive leak as the tracer gas will be quickly pumped out when test chamber is pumped down.


Helium charged sniffer test

In this last case the part is pressurized with helium. The mass spectrometer is fitted with a special device, a sniffer probe, that allows it to sample air (and tracer gas when confronted with a leak) at atmospheric pressure and to bring it into the mass spectrometer. This mode of operation is frequently used to locate a leak that has been detected by other methods, in order to allow for parts repair. Modern machines can digitally remove the helium two decades below the background level and thus it is now possible detect leaks as small as 5·10 Pa·m·s in sniffing mode.


See also

*
Mass spectrometry Mass spectrometry (MS) is an analytical technique that is used to measure the mass-to-charge ratio of ions. The results are presented as a '' mass spectrum'', a plot of intensity as a function of the mass-to-charge ratio. Mass spectrometry is u ...
* Tracer-gas leak testing *
Leak noise correlator A leak noise correlator is an electronic device used for Leak Detection and as a leak locator to find leaks in pressurized water or gas lines. Typically, microphones or acoustic sound sensors are placed in contact with the pipe, at two or more poi ...
* Helium analyzer


References


External links


Test of medical devices (FDA)

Leak Detection

UST method
{{DEFAULTSORT:Helium mass spectrometer Mass spectrometry Vacuum systems