Standard Addition Method
   HOME

TheInfoList



OR:

The method of standard addition is a type of quantitative analysis approach often used in
analytical chemistry Analytical chemistry studies and uses instruments and methods to separate, identify, and quantify matter. In practice, separation, identification or quantification may constitute the entire analysis or be combined with another method. Separati ...
whereby the standard is added directly to the aliquots of analyzed sample. This method is used in situations where sample matrix also contributes to the analytical signal, a situation known as the matrix effect, thus making it impossible to compare the analytical signal between sample and standard using the traditional
calibration curve In analytical chemistry, a calibration curve, also known as a standard curve, is a general method for determining the concentration of a substance in an unknown sample by comparing the unknown to a set of standard samples of known concentration. ...
approach.


Applications

Standard addition is frequently used in chemical
instrumental analysis Instrumental analysis is a field of analytical chemistry that investigates analytes using scientific instruments. Spectroscopy Spectroscopy measures the interaction of the molecules with electromagnetic radiation. Spectroscopy consists of many ...
such as
atomic absorption spectroscopy Atomic absorption spectroscopy (AAS) and atomic emission spectroscopy (AES) is a spectroanalytical procedure for the quantitative determination of chemical elemlight) by free atoms in the gaseous state. Atomic absorption spectroscopy is based o ...
and gas chromatography. Suppose that the concentration of silver in samples of photographic waste is to be determined by atomic-absorption spectrometry. Using the calibration curve method, an analyst could calibrate the spectrometer with some aqueous solutions of a pure silver salt and use the resulting calibration graph in the determination of the silver in the test samples. This method is only valid, however, if a pure aqueous solution of silver, and a photographic waste sample containing the same concentration of silver, give the same absorbance values. In other words, in using pure solutions to establish the calibration graph it is assumed that there are no ‘matrix effects’, i.e. no reduction or enhancement of the silver absorbance signal by other components. In many areas of analysis such an assumption is frequently invalid. Matrix effects occur even with methods such as plasma spectrometry, which have a reputation for being relatively free from interferences. The method of standard additions is usually followed to eliminate matrix effects. Experimentally, equal volumes of the sample solution are taken, all but one are separately ‘spiked’ with known and different amounts of the analyte, and all are then diluted to the same volume. The instrument signals are then determined for all these solutions and the results plotted. As usual, the signal is plotted on the y-axis; in this case the x-axis is graduated in terms of the amounts of analyte added (either as an absolute weight or as a concentration). The (unweighted) regression line is calculated in the normal way, but space is provided for it to be extrapolated to the point on the x-axis at which y = 0. This negative intercept on the x-axis corresponds to the amount of the analyte in the test sample. This value is given by b/a, the ratio of the intercept and the slope of the regression line. Similarly in gas chromatography the following procedure is used: 1) The chromatogram of the unknown is recorded 2) a known amount of the analyte(s) of interest is added 3) the sample is analyzed again under the same conditions and the chromatogram is recorded. From the increase in the peak area (or peak height), the original concentration can be computed by extrapolation. The detector response must be a linear function of analyte concentration and yield no signal (other than background) at zero concentration of the analyte.


Procedure

A typical procedure involves preparing several solutions containing the same amount of unknown, but different amounts of standard. For example, five 25 mL volumetric flasks are each filled with 10 mL of the unknown. Then the standard is added in differing amounts, such as 0, 1, 2, 3, and 4 mL. The flasks are then diluted to the mark and mixed well. The idea of this procedure is that the total concentration of the analyte is the sum of the unknown and the standard, and that the total concentration varies linearly. If the signal response is linear in this concentration range, then a plot similar to what is shown above is generated.


Error

The x-intercept gives the concentration of the unknown. Note this value is the diluted concentration. In the procedure section above, 10 mL of unknown was diluted to 25 mL. It is this diluted concentration that is found by the x-intercept. To find the original concentration of the unknown, one must back calculate that value. The error in the x-intercept is calculated as shown below. s_x=\frac\sqrt *s_y is the standard deviation in the residuals =\sqrt *m is the slope of the line *b is the y-intercept of the line *n is the number of standards *\bar is the average measurement of the standards *x_i are the concentrations of the standards *\bar is the average concentration of the standards


See also

* Standard Curve *
Isotope dilution Isotope dilution analysis is a method of determining the quantity of chemical substances. In its most simple conception, the method of isotope dilution comprises the addition of known amounts of isotopically enriched substance to the analyzed samp ...
*
Internal standard An internal standard in analytical chemistry is a chemical substance that is added in a constant amount to samples, the blank and calibration standards in a chemical analysis. This substance can then be used for calibration by plotting the ratio of ...


References

*{{cite book , last = Harris , first = Daniel C. , authorlink = , year = 2003 , title = Quantitative Chemical Analysis 6th Edition , publisher = W.H. Freeman , location = New York , id = Analytical chemistry Laboratory techniques